User's Manual

15
Samsung LABGEO
PA20
Safety Warnings and Caution
Ŷ Cautions for storage and maintenance after use
Ŷ
Do not store the product in humid areas or areas inuenced by
temperature, humidity or wind.
Ŷ
Store the device on a at surface and keep it away from vibration or impact.
Ŷ
Do not store the product with chemicals or gas.
Ŷ
Operating conditions and temperature: 10 - 32 °C (50 - 89 °F),
Humidity: 0% - 80%
Ŷ
Storage conditions and temperature: -10°C - +40°C , Humidity: 0% - 80%
Ŷ
If there is blood residue from a specimen or any contaminants in the disc
tray and the interior or exterior of the analyzer, remove the contaminant
using a soft cloth while wearing gloves.
Wet the cloth using 30% isopropyl alcohol when removing contaminants.
Do not use highly corrosive agents such as benzene or acetone.
These agents may damage the analyzer.
Ŷ
When moving or transporting the product, follow the conditions for using
and storing the product specied in this manual. Follow all cautions when
moving the product.
Ŷ
A specimen may be contaminated with pathogens or viruses that cause
infectious illnesses such as Hepatitis B. All specimens must be considered
and handled as potential contaminants.
Ŷ
Used discs must be processed using Medical Waste Treatment Laws after
use.