User's Manual
162
Samsung LABGEO
PA20
Error Code Message Problem Solution
205102B Data Transfer Error
Connection to the PC
has connection errors.
Check your network
connections.
205202B Data Transfer Error
Connection to the EMR
has connection errors.
Check your network
connections.
206001A Barcode Read Error
The barcode on the disc
is contaminated.
Try again after replacing
the contaminated disc
with a new one.
206002A Barcode Error
The disc is not for the
blood analyzer.
Replace the disc with a
disc for analysis.
206003A Barcode Error The disc has expired.
Replace the disc
with one with a valid
expiration date and run
the analysis again.
206004A
Barcode Error
Analyzer is unusable
due to disc barcode
error.
Contact Technical
Support.
206005A
206007A Barcode Error
Using disc with new
analytes requires a
software update.
Contact Technical
Support.
206008A -
206010A
Barcode Error
Unable to use the
device due to disc
barcode error.
Contact Technical
Support.
206020A Barcode Error
Unable to use the
device due to disc
barcode error.
Contact Technical
Support.
206100A-
206103A
RFID Error
Unable to use the
device due to RFID
error.
Contact Technical
Support.
207002A Analysis Info Error
Analyzer requires new
analysis information.
Contact Technical
Support.
207042C Used Disc
The disc has been used
before.
Try again after replacing
the contaminated disc
with a new one.
208202B Insucient Sample
The amount of the
injected blood sample
is insucient.
Make sure that a
sucient amount
of blood has been
injected.