User's Manual

Table Of Contents
Users Manual
http://www.exoatletasia.com
ExoAtlet
II 84 / 84 Ver 1.0 (2019.03.29.)
12.2.4. Recommended separation distance between this system and mobile RF
communication devices - Non-LIFE-SUPPORTING equipment or system
Recommended separation distances between portable and mobile RF communications equipment and the
ExoAtlet
The
ExoAtlet
is
intended
for
use
in
an
electromagnetic
environment
in
which
radiated
RF
disturbances
are controlled. The customer or the user of the ExoAtletcan help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the ExoAtletas recommended below, according to the maximum output
power of the communications equipment.
Rated maximum output power
of transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
d = 1.17
80 MHz to 800 MHz
d = 1.17
800 MHz to 2,5 GHz
d = 2.33
0.01 0.117 0.117 0.233
0.1 0.370 0.370 0.736
1 1.17 1.17 2.33
10 3.70 3.70 7.36
100 11.7 11.7 23.3
For
transmitters
rated
at
a
maximum
output
power
not
listed
above,
the
recommended
separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.