User's Manual

Samsung Confidential
[36]
Rated maximum output
power of transmitter
[W]
Separation distance according to frequency of transmitter [m]
150kHz to 80MHz
80MHz to
800MHz
800MHz to
2.5GHz
V
1
=3Vrms
E
1
=3V/m
E
1
=3V/m
0.01
0.116
0.1166
0.2333
0.1
0.368
0.3687
0.7378
1
1.166
1.1660
2.3333
10
3.687
3.6872
7.3785
100
11.660
11.6600
23.333
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where p is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
NOTE 1) At 80MHz and 800MHz, the separation distance for the higher frequency range applies.
NOTE 2) These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
6.5 Immunity and Compliance Level
Immunity test
IEC 60601 Test Level
Actual Immunity Level
Compliance Level
Conducted RF
IEC 61000-4-6
3Vrms
150kHz to 80MHz
3Vrms
3Vrms
Radiated RF
IEC 61000-4-3
3Vrms
80MHz to 2.5GHz
3V/m
3V/m