HSI-LV Isolated Low Voltage Module
Voltage input: 12 ranges (10 mV to 50 V)
Current input: ±20 mA using DAQ-SHUNT-1 (option)
±5 A using DAQ-SHUNT-4
or DAQ-SHUNT-5
Bandwidth: 2 MHz
Isolation: 1 kV RMS with banana jacks
Additional signal input types using MSI interfaces:
IEPE Constant current sensors
12 ranges (10 mV to 5 V)
RTD Resistance Temperature Detector (Pt100
to Pt2000)
9 resistance ranges (8 to 4000 Ω); requires MSI-V-RTD
CHARGE Charge up to 50000 pC requires MSI-V-CHA-50
HSI-LV Specications
Parameter HSI-LV
Input ranges ±10, ±20, ±50, ±100, ±200, ±500 mV and ±1, ±2.5, ±5, ±10, ±25, ±50 V
Push button selectable ranges ±10 mV, ±50 mV, ±200 mV, ±1 V, ±5 V, ±10 V, ±50 V
One year accuracy (1) Range Signal Frequency Accuracy
Bipolar
10 mV to 100 mV DC ±0.02 % of reading ±60 μV
2.5 V DC ±0.02 % of reading ±0.1 % of range
200 mV to 50 V DC ±0.02 % of reading ±0.05 % of range
10 mV to 100 mV
0.1 Hz to 10 kHz
>10 kHz to 50 kHz
>50 kHz to 100 kHz
>100 kHz to 1 MHz
>1 MHz to 2 MHz
±0.1 % of reading ±30 μV
±0.4 % of reading ±30 μV
±(0.016*f) % of reading ±0.1 % of range ±(0.010*f)
% of reading ±1 % of range ±(0.014*f) % of reading
±3 % of range
200 mV to 50 V
0.1 Hz to 1 kHz
>1 kHz to 10 kHz
>10 kHz to 50 kHz
>50 kHz to 100 kHz
>100 kHz to 1 MHz
>1 MHz to 2 MHz
±0.05 % of reading ±0.01 % of range ±0.1 % of
reading ±0.05 % of range ±0.4 % of reading ±0.05
% of range ±(0.016*f) % of reading ±0.1 % of range
±(0.010*f) % of reading ±1 % of range ±(0.014*f)
% of reading ±3 % of range
Unipolar
10 mV to 100 mV
200 mV to 50 V
DC
DC
±0.02 % of reading ±60 μV
±0.02 % of reading ±0.08 % of range
Input coupling DC or AC software selectable (1.5 Hz standard, custom filters upon request, down to 0.01 Hz)
Gain linearity Typically 0.01 %; max. 0.04 % of full scale
Gain drift range Typically 10 ppm/°C (max. 30 ppm/°C)
Offset drift:
10 mV to 200 mV:
500 mV to 50 V:
Typically 3 μV/°C
Typically 10 ppm of range/°C
Long term stability: 100 ppm/sqrt (1000 hrs)
Input resistance: 1 MΩ
-3dB Bandwidth: 2 MHz
Filter selection: Push button or software