When you click FILL, DS-7 will prompt you to press the black button on the EPAD2 module. Use a paperclip
or thin pencil lead to do that, and the module will be addressed to this slot number, and it will show up here
on the Setup screen. Use the SET CH. button on the far right to congure the eight channels of this module.
Now connect the next EPAD2 module and repeat these steps, moving down to the next empty amplier but-
ton, and so on, until all EPAD2 modules have been installed.
Be sure to plug the terminator into the last EPAD2 module (even if there is only one EPAD2 module).
xPAD2 Series Calibration Information
All DEWETRON modules are calibrated at 25°C after a warmup time of 30 minutes and meet their speci-
cations when leaving the factory. The time interval for recalibration depends on environmental conditions.
Typically, the calibration should be checked once a year.
Calibration certicates are available from DEWETRON as an option. DEWETRON offers several types:
NIST traceable DEWETRON calibration certicate (USA CAL LAB only)
ISO traceable DEWETRON certicate (European CAL LAB only)
Calibration certicate according to ÖKD (equivalent to DKD)
This manual contains no calibration information. There are separate resources optionally available for
MDAQ series modules for automated calibration under Met/CAL®. The CAL-KIT contains the required
cables, software and instructions that you need to add to your own calibration lab. It does not include a
calibrator or volt meter.