Operation Manual

17
inflation takes longer than 3 minutes (Er6 appears in
the display),
the cuff is too tight or too loose (Er3 or Er4 appears in
the display),
the blood-pressure values are unusually high or low
(Er1 or Er2 appears in the display),
the inflation pressure is higher than 300 mmHg (Er5
appears in the display),
there is a system or device error (Er0, Er7, Er8, Er9,
ErA or ErB appears in the display)
the batteries are almost flat; the batteries must be
replaced.
In the above cases, you must repeat the measure-
ment.
Reinsert the batteries if necessary, or else replace
them.
8. Cleaning and Care
Clean your blood pressure computer carefully using
only a slightly moistened cloth.
Do not use any detergents or solvents.
Never hold the instrument under water as otherwise
liquid can penetrate and damage the instrument.
Never place any heavy objects on the instrument.
9. Technical details
Model no. SBM 03
Measurement Oscillating, non-invasive blood
method: pressure measurement on the
wrist
Measurement range Pressure 0-300 mmHg, heart
rate 40-180 beats/minute
Precision of Systolic ± 3 mmHg /
pressure reading diastolic ± 3 mmHg
pulse ± 5% of displayed value
Measurement Max. admissible standard
uncertainty deviation according to clinical
examination: systolic 8 mmHg /
diastolic 8 mmHg
Memory 1 x 60 storable data lines
Dimensions L 84 x W 62 x H 30 mm
Weight Approx. 140 g
Cuff size For a wrist circumference of
approx. 140 to 195 mm
Admissible operating +5 °C to +40 °C, 40-80%
temperature relative humidity
Admissible -20 °C to +55 °C, 10-95%
storage temperature relative humidity
Power supply 2 x 1.5 V Micro
(Alkaline Type LR 03)
Battery life For 250 measurements
Accessories Storage box, instructions for
use, 2 x AAA batteries