Data Sheet

DIGITAL MULTIMETER
PC7000
APPLICATIONS AND FEATURES
This instrument is a portable multimeter designed for the
measurement of low-voltage circuit.
This is used at small communications equipments, home
electric appliances, voltage of lump line,various types of
batteries, and circuit analyses with additional functions.
AC True RMS / 50000 count(500000 count for DCV)
0.03% best accuracy
Dual display capability with Backlight to simultaneously display
voltage/current/frequency and AC/DC components of voltage/current
High speed bar graph
Low Pass Filter(LPF) useful for measurement of Variable Frequency Drive(VFD)
Current(mA/μA)% 4-20mA measurement
Conductance measurement / Capacitance measurement
Temperature measurement
(with K type thermocouple sensor : -50℃1000℃)
Frequency measurement(AC sine wave only)
Logic frequency measurement
Duty cycle measurement
MAX, MIN, AVE value hold / Data hold / Range hold
Relative value
Auto power save
(17min after the last operation)(cancelable)
Backlight
Protective holster with wall hanger and lead holder / Tilt stand
PC Link system with optical Link USB cable
(KB-USB7)
and software
(PC Link7)
*KB-USB7 and PC Link7 are optional accessories
*1 nS(Conductance): High-value resistance of Giga-Ohms for leakage measurements. Conductance is the inverse of Resistance, that is S=1/Ω or nS=1/GΩ
*2 Accuracy of film capacitor or equivalent with low leakage.
A battery for monitoring has been installed prior to shipment from the factory. It may be discharged before the
expiration of the described battery life. This battery is used to check the functions and performance of the
product.Specifications and external appearance of the product described above may be revised for modification
without prior notice.
Operation method
AC measuring method
Display
Range selection
Overload
Polarity
Battery low warning
Sampling rate
Power consumption
Bandwidth
EN61010-1 CAT.II
EN61010-1 CAT.III
Fuse / Battery
Size / Mass
Standard accessories included
Δ-Σ
True RMS
50000 count and 500000 count
Auto and manual ranges
"O.L" is displayed
Automatic selection(only " - " is displayed when
negative voltage is inputted.)
Battery( ) mark lights when the internal
battery's power is below approx. 7V
Numeric part : 5 times/sec(50,000 count mode),
1.25 times/sec(500,000 count mode)
Bar graph part : 60 times/sec
Approx. 48mW Approx. 0.45mW(Auto power save)
V : 40Hz3kHz, 3kHz20kHz(below 99.99V)
A : 40Hz1kHz
1000V
600V
11A/1000V IR20kA Φ10x38mm 6LR61(9V) x 1
0.4A/1000V IR30kA Φ6.3x32mm
H184 x W86 x D52mm / 430g
Test Lead(TL-23a), Holster(H-700),
Thermocouple K type (K-250PC), Instruction Manual
SPECIFICATIONS
DCV
ACV
DCA
ACA
Resistance
Capacitance
Temperature
Frequency
Logic frequency
Duty cycle
dBm
Continuity
Diode test
Measuring range
500m/5/50/500/1000V
500m/5/50/500/1000V
500μ/5000μ/50m/500m/5/10A
500μ/5000μ/50m/500m/5/10A
500/5k/50k/500k/5M/50MΩ/99.99nS *
1
50n/500n/5μ/50μ/500μ/5m/25mF
-50℃1000℃(thermocouple K type)
10Hz200kHz
5Hz2MHz
0.1%99.99%
-29.83dBm54.25dBm
Buzzer sounds at between 20Ω and 200Ω Open voltage: approx. 1.3V
Open voltage: approx. 3V
Best accuracy
±(0.03%+2)
±(0.5%+40)
±(0.1%+20)
±(0.6%+40)
±(0.2%+6)
±(0.8%+3)*
2
±(0.3%+2)
±(0.02%+4)
±(0.002%+4)
±(3d/kHz+2)
±(0.25dB+2)
Resolution
0.01mV
0.01mV
0.01μA
0.01μA
0.01Ω
0.01nF
0.1℃
0.001Hz
0.001Hz
0.01%
0.01dB
Input impedance
10MΩ