INSTRUCTION MANUAL MOV-112S MOV-212S Sterilizer Библиотека Ладовед. OCR Юрий Войкин 2008г.
Note: 1. No part of this manual may be reproduced in any form without the expressed written permission of SANYO. 2. The contents of this manual are subject to change without notice. 3. Please contact SANYO if any point in this manual is unclear or if there are any inaccuracies. SANYO Electric Biomedical Co., Ltd. All rights reserved. Printed in Japan. CONTENTS PRECAUTIONS FOR SAFE OPERATION P. 2 CAUTIONS FOR USAGE P. 6 ENVIRONMENTAL CONDITIONS P. 6 STERILIZER COMPONENTS P. 7 INSTALLATION P.
PRECAUTIONS FOR SAFE OPERATION It is imperative that the user complies with this manual as it contains important safety advice. Items and procedures are described so that you can use this unit correctly and safely. If the precautions advised are followed, this will prevent possible injury to the user and any other person. Precautions are illustrated in the following way: WARNING Failure to observe WARNING signs could result in a hazard to personnel possibly resulting in serious injury or death.
PRECAUTIONS FOR SAFE OPERATION WARNING Do not use the unit outdoors. rain water. Current leakage or electric shock may result if the unit is exposed to Only qualified engineers or service personnel should install the unit. The installation by unqualified personnel may cause electric shock or fire. Install the unit on a sturdy floor. If the floor is not strong enough or the installation site is not adequate, this may result in injury from the unit falling or tipping over.
PRECAUTIONS FOR SAFE OPERATION !\WARNING Ensure you do not inhale or consume medication or aerosols from around the unit at the time of maintenance. These may be harmful to your health. Never splash water directly onto the unit as this may cause electric shock or short circuit. Never disassemble, repair, or modify the unit yourself. Any such work carried out by an unauthorized person may result in fire or injury due to a malfunction.
PRECAUTIONS FOR SAFE OPERATION CAUTION Select a level and sturdy floor for installation. This precaution will prevent the unit from tipping. Improper installation may result in water spillage or injury from the unit tipping over. Connect the unit to a power source as indicated on the rating label attached to the unit. Use of any other voltage or frequency other than that on the rating label may cause fire or electric shock. tfB Fix the shelves securely. Incomplete installation may cause injury or damage.
CAUTIONS FOR USAGE 1. This unit is not explode-proof construction. Never store the materials which cause flammable or volatile gas and never store the flammable or volatile liquid in the unit. 2. The door glass, inside door panel, and exhaust vent are very hot when the unit is used with high temperature setting. Never touch such portion directly with the hands. 3. Avoid to open the door when the chamber temperature is more than 100°C as it is dangerous.
STERILIZER COMPONENTS
STERILIZER COMPONENTS 1. Door: During use, inner panel of the door becomes extremely hot; Be careful. 2. Observation window: 3. Leveling foot: 4. Fan (inside): 5. Control panel: During use, glass and its surrounding become very hot; Be careful. Made of rubber. Propeller type. Can be adjusted by screws. To cool down the heat discharge plate and fan motor. Refer to page 9. 6. Separation plate: Do not place objects directly on the plate.
STERILIZER COMPONENTS Control panel and keypad 1 2 3 1. Power switch (POWER): 4 5 6 7 8 Power switch of the control panel. 2. Alarm buzzer stop key (BZ): Press this key to silence the buzzer in the event that the alarm operates and the buzzer sounds. Press it once again to reactive the buzzer. 3. Timer mode select key (TIMER MODE): By pressing this key, the timer mode is selected. Refer to the "Timer function" on page 13. 4.
INSTALLATION Installation site To operate this unit properly and to obtain maximum performance, install the unit in a location with the following conditions: 1. A location not subjected to direct sunlight or direct air flow from an air conditioner 2. A location with adequate ventilation Leave at least 30 cm around the unit for ventilation. performance. 3.
INSTALLATION Installation 1. Remove the packaging materials and tapes Remove all transportation packaging materials and tapes. Open the doors and ventilate the unit. If the outside panels are dirty, clean them with a neutral detergent and wipe it up with a wet cloth. 2. Adjust the leveling feet Extend the leveling legs by rotating them counterclockwise so they contact the floor or bench. Ensure the unit is level. 3.
OPERATING INSTRUCTIONS Basic operating flow chart of control panel 1. Start of operation when the input of setting is conducted.
OPERATING INSTRUCTIONS Timer function The timer function has two modes as follows: 1. Normal mode: starts immediately after start of the running 2. Timer mode 1: starts when the chamber temperature is 2.0°C lower than the set point. When conducting the sterilization, always select "Timer mode 1". On timer mode 1, the timer starts when the chamber temperature is 2.0°C lower than the set point.
OPERATING INSTRUCTIONS Stable running Table below shows the basic procedure for setting the chamber temperature. the sequence indicated in the table. Perform key operations in The example in the table is based on the assumption that the desired temperature is 50°C and continuous running. 50°C Time Basic operation sequence (Example: Chamber temperature; 50 C, Continuous running) Description of operation 1 Press the power key.
OPERATING INSTRUCTIONS Stable running with delay time 50°C Delay time Basic operation sequence (Example: Chamber temperature; 50°C, Delay time; 1 hour) Description of operation Key operated Indication after operation Press the power key. POWER О ALARM О SET VALUE О MODE 1 О DELAY О START E5\ (The current chamber temperature is displayed.) (Delay time reset) Press the delay timer key. Set the time to 01:00 with the digit shift key and the numeric value shift key.
OPERATING INSTRUCTIONS Sterilization with timer mode 1 200°C 2 hours Basic operation sequence (Example: Description of operation Press the power key. Chamber temperature; 180°C, Delay time; 2 Ilours) Key operated Indication after operation POWER 0 ALARM 0 SET VALUE О MODE 1 DELAY О ALARM • SET VALUE 0 MODE 1 О DELAY О START i 1 i 1 E5 1 (The current chamber temperature is displayed.) 2 3 4 (Release of delay set time) Press the delay timer key.
ALARMS & SAFETY FUNCTIONS This unit has the alarms and safety functions shown in table below, and also self diagnostic functions. Alarms and safety functions Alarm & Indication Situation Safety If the internal temperature Automatic set deviates from the set Alarm lamp lights. temperature temperature by +10°C or more All digits on the Alarm (and if timer mode 1, by -5°C or indicator flash. more) If a key operation is not Auto return performed for about 45 sec. in Normal display mode. each setting mode.
ROUTINE MAINTENANCE LWARNING Always disconnect the power supply to the unit prior to any repair or maintenance of the unit in order to prevent electric shock or injury. Ensure you do not inhale or consume medication or aerosols from around the unit at the time of maintenance. These may be harmful to your health. CAUTION Always put on dry gloves to protect hands at the time of maintenance. Failure to use gloves may result in cuts or abrasions from any sharp edges or corners.
TROUBLE SHOOTING If the unit malfunctions, check out the following before calling for service. The unit does not operate at all 1. The unit is not plugged correctly into a power outlet. 2. The circuit breaker at the power source is active. 3. A power failure has occurred. 4. A fuse has blown. The key operation is disabled 1. The key lock function is set in OFF mode. If the alarm function operates If the alarm function and the buzzer operates, check the cause using the following procedure.
SPECIFICATIONS Sterilizer Name Model MOV-112S MOV-212S External dimensions W580 x D595 x H820 (mm) W730 x D645 x H870 (mm) Internal dimensions W450 x D450x H450 (mm) W600 x D500x H500 (mm) 90 L 150 L Effective capacity Exterior Electrically galvanized steel plate, Acrylic resin painted Interior Stainless steel plate (SUS 304) Electrically galvanized steel plate, Acrylic resin painted Door Observation window Reinforced glass x 3, W150 x H380 mm Glass wool Insulation Rack Stainless steel pl
CAUTION Please fill in this form before servicing. Hand over this form to the service engineer to keep for his and your safety. Safety check sheet 1. Refrigerator contents : Risk of infection: Risk of toxicity: Risk from radioactive sources: •Yes •Yes •Yes •Yes • • • • No No No No (List all potentially hazardous materials that have been stored in this unit.) Notes : 2.