CONTENTS INTRODUCTION P. 2 PRECAUTIONS FOR SAFE OPERATION P. 3 ENVIRONMENTAL CONDITIONS P. 7 FREEZER COMPONENTS P. 8 Control panel P. 10 INSTALLATION SITE P. 11 INSTALLATION P. 12 START-UP OF UNIT P. 12 TEMPERATURE SETTING P. 13 KEY LOCK FUNCTION P. 13 REMOTE ALARM TERMINAL P. 14 ALARM TEMPERATURE SETTING P. 15 ALARMS & SAFETY FUNCTIONS P. 16 DEFROSTING P. 17 MAINTENANCE P. 17 REPLACEMENT OF BATTERY P. 18 TROUBLESHOOTING P. 19 DISPOSAL OF UNIT P.
INTRODUCTION ■ Read this manual carefully before using the appliance and follow the instructions for safety operation. ■ Sanyo never guarantee any safety if the appliance is used for any objects other than intended use or used by any procedures other than those mentioned in this manual. ■ Keep this manual in an adequate place to refer to it as necessary. ■ The contents of the manual will be subjected to change without notice due to the improvement of performance or functions.
PRECAUTIONS FOR SAFE OPERATION It is imperative that the user complies with this manual as it contains important safety advice. Items and procedures are described so that you can use this unit correctly and safely. If the precautions advised are followed, this will prevent possible injury to the user and any other person. Precautions are illustrated in the following way: WARNING Failure to observe WARNING signs could result in a hazard to personnel possibly resulting in serious injury or death.
PRECAUTIONS FOR SAFE OPERATION WARNING Do not use the unit outdoors. Current leakage or electric shock may result if the unit is exposed to rain water. Only qualified engineers or service personnel should install the unit. unqualified personnel may cause electric shock or fire. The installation by Install the unit on a sturdy floor and take an adequate precaution to prevent the unit from turning over.
PRECAUTIONS FOR SAFE OPERATION WARNING Ensure you do not inhale or consume medication or aerosols from around the unit at the time of maintenance. These may be harmful to your health. Never splash water directly onto the unit as this may cause electric shock or short circuit. Never put containers with liquid on the unit as this may cause electric shock or short circuit when the liquid is spilled. Never bind, process, or step on the power supply cord, or never damage or break the power supply plug.
PRECAUTIONS FOR SAFE OPERATION CAUTION Use a dedicated power source (a dedicated circuit with a breaker) as indicated on the rating label attached to the unit. A branched circuit may cause fire resulting from abnormal heating. Connect the power supply plug to the power source firmly after removing the dust on the plug. A dusty plug or improper insertion may cause a heat or ignition. Never store corrosive substances such as acid or alkali in this unit if the container cannot be sealed.
ENVIRONMENTAL CONDITIONS This equipment is designed to be safe at least under the following conditions (based on the IEC-61010-1): ■ Indoor use; ■ Altitude up to 2000 m; ■ Ambient temperature 5oC to 40oC ■ Maximum relative humidity 80% for temperature up to 31oC decreasing linearly to 50% relative humidity at 40oC; ■ Mains supply voltage fluctuations not to exceed ±10% of the nominal voltage; ■ Other supply voltage fluctuations as stated by the manufacturer; ■ Transient overvoltages according to Installatio
FREEZER COMPONENTS 12 14 13 1 2 11 10 3 9 8 7 6 4 5 15 16 17 18 8
FREEZER COMPONENTS 1. Lock: Turn clockwise to 180o with a key and the outer door is securely locked. 2. Magnetic door gasket: This prevent the cold air from escaping. Always keep clean. 3. Handle: Always grip this handle to open and close the outer door. 4. Drain hose: The drain hose can be used to collect the defrosted water. See page 17 for usage. 5. Leveling foot: The height of the freezer can be adjusted by this screw type foot. level at the installation. Keep the unit in 6.
FREEZER COMPONENTS Control panel 7 8 9 DOOR 1 2 3 4 1. Alarm test key (ALARM TEST): 5 6 Buzzer will sound and remote alarm will be activated by pressing. 2. Alarm buzzer stop key (BUZZER): Press this key to silence the buzzer in the event that the alarm operates and buzzer sounds. See page 16 for the details. 3. Defrost key (DEF): Defrost will be started by pressing this key for more than 5seconds. Pressing this key again after defrosting leads resumption of the operation.
INSTALLATION SITE To operate this unit properly and to obtain maximum performance, install the unit in a location with the following conditions: ■ A location not subjected to direct sunlight Do not install the unit under direct sunlight. Installation in a location subjected to direct sunlight cannot obtain the intended performance. ■ A location with adequate ventilation Leave at least 10 cm around the unit for ventilation.
INSTALLATION 1. Removing the packaging materials and tapes Remove all transportation packaging materials and tapes. Open the doors and ventilate the unit. If the outside panels are dirty, clean them with a diluted neutral dishwashing detergent. (Undiluted detergent can damage the plastic components. For the dilution, refer to the instruction of the detergent.) After the cleaning with the diluted detergent, always wipe it off with a wet cloth. Then wipe off the panels with a dry cloth. Leveling foot 2.
TEMPERATURE SETTING Table 1 shows the basic operation method. Perform key operations in the sequence indicated in the table. The example in the table is based on the assumption that the freezer temperature is -20oC. Note: The unit is set at the factory that the chamber temperature -30oC. Table 1 Basic operation sequence (Example: Chamber temperature -25oC) Description of operation Key operated Indication after operation The current chamber temperature is Connect the power source and turn displayed.
REMOTE ALARM TERMINAL WARNING Always disconnect the power supply cord before connecting an alarm device to the remote alarm terminal. The terminal of the remote alarm is installed at the top of the unit. The alarm is outputted from this terminal between terminal 1 and terminal 2. Contact capacity is DC 30 V, 2 A. between COM. and N.O. between COM. and N.C. At normal Open Close At abnormal Close Open 1 2 3 Battery switch N.C. COM N.O.
ALARM TEMPERATURE SETTING This unit is provided with both high and low temperature alarms. The temperature at which the alarm is activated may be changed. The available set range for high temperature alarm is between +5oC and +15oC, and -5oC and -15oC for low temperature alarm against the chamber temperature. Note: The temperature alarm is set at ±10oC of the set temperature at the factory. Table 3.
ALARMS & SAFETY FUNCTIONS This unit has the alarms and safety functions shown in Table 4, and also self diagnostic functions. Table 4 Alarms and safety functions Alarm & Safety High temperature alarm Low temperature alarm Situation If the chamber temperature o deviates by +10 C (factory set, changeable) from setting. If the chamber temperature deviates by -10oC (factory set, changeable) from setting. Indication Alarm lamp is flashed. Temperature indicator is flashed. Alarm lamp is flashed.
DEFROSTING Use the scraper provided for removing the frost if the freezer operation must be continued. Pay attention not to impact or damage the inner wall. 1. When defrosting, move all the contents of containers in the freezer to another low temperature freezer. 2. Pressing the DEF key for more than 5 seconds to stop the refrigerating operation. While the refrigerating operation is stopped, the current chamber temperature and dF is displayed on the control panel alternately. 3.
REPLACEMENT OF BATTERY Location of a nickel-metal-hydride battery This unit is provided a nickel-metal-hydride battery for the power failure warning device. The battery is located in the battery mounting plate inside the cover on the upper side. (Fig. 1) The high voltage components are enclosed in the front cover. The cover should be removed by a qualified engineer or a service personnel only to prevent the electric shock. Removal of nickel-metal-hydride battery 1.
TROUBLESHOOTING If the unit malfunctions, check out the following before calling for service. Malfunction The chamber is not cooled at all ■ ■ ■ ■ Check/Remedy The circuit breaker of power source is active. The voltage is too low? (In this case, call an electrician.) The power switch is not ON. The large amount of articles (load) is stored in the chamber at one time. ■ The freezer is in defrost condition.
DISPOSAL OF UNIT WARNING If the unit is to be stored unused in an unsupervised area for an extended period ensure that children do not have access and doors cannot be closed completely. The disposal of the unit should be accomplished by appropriate personnel. Always remove doors to prevent accidents such as suffocation. Recycle of battery The unit contains a rechargeable battery. The battery is recyclable. At the end of it’s useful life, check with you local solid officials option or proper disposal.
DISPOSAL OF UNIT Note: This symbol mark and recycle system are applied only to EU countries and not applied to the countries in the other area of the world. Waste Electrical and Electronic Equipment (WEEE) Directive-2002/96/EC (English) Your SANYO product is designed and manufactured with high quality materials and components which can be recycled and reused. This symbol means that electrical and electronic equipment, at their end-of-life, should be disposed of separately from your household waste.
DISPOSAL OF UNIT (French) Votre produit Sanyo est conçu et fabriqué avec des matèriels et des composants de qualité supérieure qui peuvent être recyclés et réutilisés. Ce symbole signifie que les équipements électriques et électroniques en fin de vie doivent être éliminés séparément des ordures ménagères. Nous vous prions donc de confier cet équipement à votre centre local de collecte/recyclage.
DISPOSAL OF UNIT (Portuguese) O seu produto SANYO foi concebido e produzido com materiais e componentes de alta qualidade que podem ser reciclados e reutilizados. Este símbolo significa que o equipamento eléctrico e electrónico no final da sua vida útil deverá ser descartado separadamente do seu lixo doméstico. Por favor, entregue este equipamento no seu ponto local de recolha/reciclagem. Na União Europeia existem sistemas de recolha separados para produtos eléctricos e electrónicos usados.
DISPOSAL OF UNIT (Dutch) Sanyo producten zijn ontwikkeld en gefabriceerd uit eerste kwaliteit materialen, de onderdelen kunnen worden gerecycled en weer worden gebruikt. Het symbool betekent dat de elektrische en elektronische onderdelen wanneer deze vernietigd gaan worden , dit separaat gebeurt van het normale huisafval. Zorg ervoor dat het verwijderen van de apparatuur bij de lokaal erkende instanties gaat gebeuren.
TEMPERATURE RECORDER (OPTION) WARNING Always disconnect the power supply to the unit prior to attachment of a temperature recorder in order to prevent electric shock or injury. An automatic temperature recorders is available for the freezer as the optional component. The type of the recorder is MTR-G85 and MTR-4015LH. For the attachment, the mounting kit is necessary. 1. Remove the screw on the cover for the recorder space on the upper front of the freezer. Fig. 1 Cover Shaft Screw 2.
TEMPERATURE RECORDER (OPTION) 5. Remove 2 rubber caps (outside and inside) on the access port on the back of the freezer and also remove the insulation in the port. Fig. 4 Access port Fig. 4 6. A recorder sensor is to do fix with supplied nylon clips the inside wall by screw. See fig. 5. Clip (Big) Fig. 5 7. Replace the insulation into the access port and cover the port completely with the rubber caps. Seal the access port with a silicon See fig. 6. 8.
TEMPERATURE RECORDER (OPTION) 1. Attach the recorder to the mounting kit (MTR-S7) by following the procedure shown in the manual enclosed with the mounting kit. Cover Shaft 2. Remove the screw on the cover for the recorder space on the upper front of the freezer. Open the cover. Fig. 7 3. Join the wire in the unit compartment with the recorder wire by each wire connector. See Fig. 8 4.
SPECIFICATIONS Name Biomedical Freezer Model MDF-U730M External dimensions W770 x D830 x H1955 (mm) Internal dimensions W650 x D710 x H1520 (mm) Effective capacity 690 L Exterior Painted steel Interior Painted steel Door Painted steel Lock 1 Insulation Shelf Rigid polyurethane foamed-in place Polyethylene coated wire 4 pieces Size: W626 x D620 (mm) Max.
PERFORMANCE Cooling performance Noise level -20oC to -30oC (ambient temperature; 5oC to 35oC, no load) 42 dB [A] (background noise; 20 dB) Maximum pressure 2.55 MPa Rated voltage AC 115 V Rated frequency 60 Hz Rated power consumption 260 W Note : • Specifications will be subject to change without notice.
CAUTION Please fill in this form before servicing. Hand over this form to the service engineer to keep for his and your safety. Safety check sheet 1. Freezer contents : Risk of infection: Risk of toxicity: Risk from radioactive sources: □Yes □Yes □Yes □Yes □No □No □No □No (List all potentially hazardous materials that have been stored in this unit.) Notes : 2. Contamination of the unit: Unit interior: No contamination: Decontaminated: Contaminated: Others: □Yes □Yes □Yes □Yes □No □No □No □No 3.
0326501002900-3 (01 August 2008) 5-5, Keihan-Hondori 2-Chome Moriguchi City, Osaka 570-8677 Japan Printed in Taiwan