User's Manual

26
TEMPERATURE RECORDER (OPTION)
5. Remove 2 rubber caps (outside and inside) on the
access port on the back of the freezer and also
remove the insulation in the port. Fig. 4
6. A recorder sensor is to do fix with supplied nylon
clips the inside wall by screw. See fig. 5.
7. Replace the insulation into the access port and
cover the port completely with the rubber caps. Seal
the access port with a silicon See fig. 6.
8. Operate the freezer until the chamber temperature
gets to the set temperature. Check the recorded
temperature and chamber temperature displayed on
the control panel. Adjust the zero adjustment volume
on the temperature recorder so that the recorded
temperature can corresponds with the displayed
temperature if they are not compliance each other
Fig. 6
Fig. 4
Fig. 5
Clip (Big)
Access port