CONTENTS INTRODUCTION P. 2 PRECAUTIONS FOR SAFE OPERATION P. 3 ENVIRONMENTAL CONDITIONS P. 7 REFRIGERATOR COMPONENTS P. 8 Control panel P. 10 Inside of front cover P. 11 INSTALLATION SITE P. 12 INSTALLATION P. 13 START-UP OF UNIT P. 14 STOCK OF CONTAINERS P. 15 CHAMBER TEMPERATURE SETTING P. 16 ALARM TEMPERATURE SETTING P. 17 SETTING OF DELAY OF DOOR ALARM P. 18 SETTING OF ALARM RESUME TIME P. 19 KEY LOCK FUNCTION P. 20 DEFROST CYCLES P. 21 REMOTE ALARM TERMINAL P.
INTRODUCTION Ŷ Read this manual carefully before using the appliance and follow the instructions for safety operation. Ŷ Sanyo never guarantee any safety if the appliance is used for any objects other than intended use or used by any procedures other than those mentioned in this manual. Ŷ Keep this manual in an adequate place to refer to it as necessary. Ŷ The contents of the manual will be subjected to change without notice due to the improvement of performance or functions.
PRECAUTIONS FOR SAFE OPERATION It is imperative that the user complies with this manual as it contains important safety advice. Items and procedures are described so that you can use this unit correctly and safely. If the precautions advised are followed, this will prevent possible injury to the user and any other person. Precautions are illustrated in the following way: WARNING Failure to observe WARNING signs could result in a hazard to personnel possibly resulting in serious injury or death.
PRECAUTIONS FOR SAFE OPERATION WARNING Do not use the unit outdoors. Current leakage or electric shock may result if the unit is exposed to rain water. Only qualified engineers or service personnel should install the unit. unqualified personnel may cause electric shock or fire. The installation by Install the unit on a sturdy floor and take an adequate precaution to prevent the unit from turning over.
PRECAUTIONS FOR SAFE OPERATION WARNING Ensure you do not inhale or consume medication or aerosols from around the unit at the time of maintenance. These may be harmful to your health. Never splash water directly onto the unit as this may cause electric shock or short circuit. Never put containers with liquid on the unit as this may cause electric shock or short circuit when the liquid is spilled. Never bind, process, or step on the power supply cord, or never damage or break the power supply plug.
PRECAUTIONS FOR SAFE OPERATION CAUTION Use a dedicated power source (a dedicated circuit with a breaker) as indicated on the rating label attached to the unit. A branched circuit may cause fire resulting from abnormal heating. Connect the power supply plug to the power source firmly after removing the dust on the plug. A dusty plug or improper insertion may cause a heat or ignition. Never store corrosive substances such as acid or alkali in this unit if the container cannot be sealed.
ENVIRONMENTAL CONDITIONS This equipment is designed to be safe at least under the following conditions (based on the IEC 61010-1): 1. Indoor use; 2. Altitude up to 2000 m; 3. Ambient temperature 5oC to 40oC 4. Maximum relative humidity 80% for temperature up to 31oC decreasing linearly to 50% relative humidity at 40oC; 5. Mains supply voltage fluctuations not to exceed ±10% of the nominal voltage; 6. Other supply voltage fluctuations as stated by the manufacture; 7.
REFRIGERATOR COMPONENTS 2 1 3 4 16 5 15 6 7 14 8 13 12 11 10 9 MPR-1411 17 12 18 MPR-1411R Rear side - 46 - 8
REFRIGERATOR COMPONENTS 1. Front cover: A front cover opens when a bottom part is pulled to the front. Open this cover when you use cleaning of the condenser filter, a remote alarm terminal. Refer to page 11. 2. Control panel: 3. Light switch: Panel opens when the right top of the lower part cover is pushed. Refer to page 10. This switch is used for turning the fluorescent lamp off and on. 4. Space for temperature recorder: An temperature recorder (optional accessory) can be mounted here. See page 32. 5.
REFRIGERATOR COMPONENTS Control panel 1 2 3 7 6 5 4 1. Door check indicator (DOOR): The red LED lamp is lit when the outer door is opened. 2. Alarm lamp (ALARM): This lamp is flashed when the audible alarm is activated. 3. Digital temperature indicator: temperature. This indicator shows the present chamber temperature or set 4. Numerical value shift key ( ): Pressing this key in the setting mode causes the numerical value to shift.
REFRIGERATOR COMPONENTS Inside of front cover 2 1 3 4 5 6 1. Condenser filter: This filter is for prevention of condenser clogged up. Clean the filter every 3 month. 2. Remote alarm terminal: This is used to connect the unit to an exterior alarm to notify users of any malfunction. Refer to page 21. 3. Glow starter: This is for the fluorescent lamp. It is recommended that the glow starter is also replaced when the fluorescent lamp is replaced. Refer to page 25. 4.
INSTALLATION SITE To operate this unit properly and to obtain maximum performance, install the unit in a location with the following conditions: Ŷ A location not compliance with the following conditions may cause poor performance, failure or accident. Ŷ A location not subjected to direct sunlight Do not install the unit under direct sunlight. Installation in a location subjected to direct sunlight may degrade the performance.
INSTALLATION 1. Removing the packaging materials and tapes Remove all transportation packaging materials and tapes. Open the doors and ventilate the unit. If the outside panels are dirty, clean them with a diluted neutral dishwashing detergent. (Undiluted detergent can damage the plastic components. For the dilution, refer to the instruction of the detergent.) After the cleaning with the diluted detergent, always wipe it off with a wet cloth. Then wipe off the panels with a dry cloth. 2.
START-UP OF UNIT Follow the procedures for the initial and consequent operations of the unit. 1. Connect the power cord to the dedicated outlet with appropriate rating. CAUTION If the unit is unplugged or the power to the unit is interrupted, do not restart the unit for at least 5 minutes. This protects the compressor. 2. On start-up, the alarm buzzer sometimes operates. In this case, stop the buzzer by pressing the alarm buzzer stop key (BUZZER). 3. Set the chamber temperature to 5oC. 4.
STOCK OF CONTAINERS Always distribute items so as not to disturb the air circulation in the chamber. Disruption of the air flow can cause items to freeze or reduce the uniformity of the chamber temperature. Never put any articles on the top of the unit. Intake air vent Do not block this vent with containers. If this vent becomes blocked, temperature regulation becomes unstable. Do not place paper or vinyl near the vent as they may be sucked into the fan.
CHAMBER TEMPERATURE SETTING Table 1 shows the basic procedure for setting the chamber temperature. Perform key operations in the sequence indicated in the table. The example in the table is based on the assumption that the desired temperature is 10oC. Note: The unit is set at the factory with a chamber temperature of 5oC. Table 1 Basic operation sequence (Example: Chamber set temperature 10oC) Description of operation Key operated Indication after operation 1 Connect the power supply plug.
ALARM TEMPERATURE SETTING This unit is provided with both high and low temperature alarms. The temperature at which the alarm is activated may be changed. The available set range for high temperature alarm is between +2oC and +14oC and -2oC and -14oC for low temperature alarm against the chamber temperature. Note: The temperature alarm is set at r5oC of the set temperature at the factory.
SETTING OF DELAY OF DOOR ALARM The door check indicator is light when the door is opened, and the alarm buzzer sounds with some delay to notice the door opening. The delay time (between lighting of the door check indicator and activation of the alarm buzzer) can be changed. Set an appropriate delay time according to the condition of use to prevent the rise of chamber temperature resulting from inadequate door close. (The available set range of delay time is between 1 and 15 minutes.
SETTING OF ALARM RESUME TIME The alarm buzzer and remote alarm are silenced by pressing he alarm buzzer stop key (BUZZER) on the control panel during alarm condition. The buzzer and remote alarm will be activated again after certain suspension if the alarm condition is continued. The suspension time can be set by following the procedure shown in the Table 5 below. The example in the table is based on the assumption that the desired duration is 20 minutes.
KEY LOCK FUNCTION This unit is provided with a key lock function. When the key lock is ON, change of temperature setting through the key pad is not available. The key lock is set in OFF at the factory.
DEFROST CYCLES 㩵 Cycle defrost When the ambient humidity is high, or a large amount of damp product is being stored inside the chamber, there is a possibility that cycle defrost may not be enough to remove all of the frost on the evaporator. In this case, a forced defrost cycle can be initiated. 㩵 Forced defrost When the unit is operating under a forced defrost cycle, the current chamber temperature and dF is displayed alternately on the digital temperature display.
ALARMS & SAFETY FUNCTIONS This unit has the alarms and safety functions shown in Table 7, and also self diagnostic functions. Table 7 Alarms and safety functions Alarm & Safety High temperature alarm Low temperature alarm Situation Indication If the chamber temperature deviates from the Alarm lamp is flashed. set temperature +2oC or up to +14oC. Chamber temperature is flashed. If the chamber temperature deviates from the set temperature -2oC or up to -14oC.
ROUTINE MAINTENANCE WARNING Always disconnect the power supply to the unit prior to any repair or maintenance of the unit in order to prevent electric shock or injury. Ensure you do not inhale or consume medication or aerosols from around the unit at the time of maintenance. These may be harmful to your health. CAUTION Always wear dry gloves to protect hands at the time of maintenance. Failure to wear gloves may result in injury from edges and corners. Cleaning of cabinet 㩵 Clean the unit once a month.
ROUTINE MAINTENANCE Replacement of fluorescent lamp The fluorescent lamp is placed vertically at the center of the chamber. Follow the procedure below to replace the lamp. 1. Turn off the fluorescent light switch and disconnect the power supply plug. Lamp 2. Remove all shelves or drawers. Fig. 3. In the case of MPR-1411R, remove the rail that is placed vertically at the center of the frame. Connector water- 4. Pull the lamp toward the rear together with the cover as shown in Fig.
ROUTINE MAINTENANCE Replacement of glow starter A glow starter is located inside the front cover. Water-proof rubber 1. Disconnect the power supply plug. Socket 2. Remove the water-proof rubber. 3. Remove the glow starter and replace with a new one. 4. Replace the water-proof rubber. Cleaning of condenser filter 1. Lift the front cover as shown in the figure. 2. Take out the shaft and insert the end of the shaft in the black hole of the front cover. Front 3.
TROUBLE SHOOTING If the unit malfunctions, check out the following before calling for service. Malfunction If nothing operates even when switched on Check/Remedy 㩵 There is a power failure. 㩵 The circuit breaker is activated. 㩵 The unit is not connected to the power supply. When no key operation is available 㩵 The key lock is set in ON (L 1). The alarm device is activated < On start-up > 㩵 The temperature in the unit does not match the set value.
DISPOSAL OF UNIT WARNING If the unit is to be stored unused in an unsupervised area for an extended period ensure that children do not have access and doors cannot be closed completely. The disposal of the unit should be accomplished by appropriate personnel. Always remove doors to prevent accidents such as suffocation.
DISPOSAL OF UNIT (English) FOR EU USERS The symbol mark and recycling systems described below apply to EU countries and do not apply to countries in other areas of the world. Your SANYO product is designed and manufactured with high quality materials and components which can be recycled and/or reused. The symbol mark means that electrical and electronic equipment, batteries and accumulators, at their end-of-life, should be disposed of separately from your household waste.
DISPOSAL OF UNIT (French) POUR LES UTILISATEURS DE UE Le symbole et les systèmes de recyclage évoqués ci-dessous s'appliquent uniquement aux pays de UE. Votre produit SANYO est conçu et fabriqué avec des composants et des matériaux de hautes qualités qui peuvent être recyclés et/ou réutilisés. Le symbole signifie que les équipements électriques et électroniques, les batteries et les accumulateurs ne doivent pas être mis au rebut avec les déchets domestiques à l'issue de leur durée de vie.
DISPOSAL OF UNIT (Portuguese) PARA UTILIZADORES DA UE O símbolo e os sistemas de reciclagem descritos abaixo aplicam-se aos países da UE e não se aplicam aos países noutras áreas do mundo. O seu produto SANYO foi concebido e fabricado com materiais e componentes de elevada qualidade que podem ser reciclados e/ou reutilizados. O símbolo significa que o equipamento eléctrico e electrónico, baterias e acumuladores, em final de vida, não devem ser deitados fora juntamente com o lixo doméstico.
DISPOSAL OF UNIT (Dutch) VOOR GEBRUIKERS IN DE EU Het symbool en de recycleersystemen die hieronder beschreven worden, zijn van toepassing op de landen in de EU en zijn niet van toepassing op landen in andere delen van de wereld. Uw SANYO product is ontworpen en gemaakt met materialen en onderdelen van hoge kwaliteit, die gerecycleerd en opnieuw gebruikt kunnen worden.
OPTIONAL COMPONENTS Automatic temperature recorder An automatic temperature recorder is available for this refrigerator as an optional accessory. If an automatic temperature recorder is required, contact your Sanyo dealer. Two recorders are selectable for the refrigerator; MTR-0621LH and MTR-G04. Please note that each recorder needs a exclusive fixture for mounting. See the following for the details.
OPTIONAL COMPONENTS 6. An attached binder (two pcs) is used for the fan cover of the top of the chamber as the Fig. 5, and a temperature sensor is fixed. 7. Route the capillary on the top of the chamber and fix it with the provided clip and screw. (Fig. 6) 8. Make a cut on the rubber caps so that the capillary can pass the caps and place them back on the access port (inside and outside). (Fig. 7) Binder Fig. 5 Clip Capillary Fig. 6 Rubber cap Capillary Fig. 7 Attachment of recorder MTR-G04 1.
OPTIONAL COMPONENTS 3. Remove the black rubber caps (inside and outside) on the access port located at the top corner of the unit and take the insulation out of the access port. Temperature recorder sensor 4. Pass the recorder sensor into the chamber through the access port. (Fig. 3) 5. An attached binder (two pcs) is used for the fan cover of the top of the chamber as the Fig. 4, and a temperature sensor is fixed. Fig. 3 4.
OPTIONAL COMPONENTS Attachment of battery kit MPR-48B The alarm lamp blinks and the alarm buzzer sounds to notice the power failure when a battery for power failure alarm is installed. For the installation of the battery for power failure alarm, a battery kit (MPR-48B), an optional component is necessary. A battery for power failure alarm is included in the battery kit. 㩵 Contact Sanyo sales representative or agent for the installation of the battery kit (MPR-48B). 1.
OPTIONAL COMPONENTS 5. Put the battery fixture over the battery and fix the battery box and battery fixture by four screws. (Fig. 4) Battery fixture Clip 6. Each harness is connected to the battery connectors (Fig. 5) and switches to a fixed mounting plate on the battery clip.(Fig. 4) 6. Connect the harness to the battery connector and battery switch (Fig. 6) and bind the harnesses by the clip on the battery fixture. Screw Screw 7.
OPTIONAL COMPONENTS Setting for battery before starting Always perform the setting shown below after the installation of battery for power failure alarm, an optional component (MPR-48B). The chamber temperature and “E09” is displayed on the temperature display alternately if the battery switch is turned on without following procedure. 㩵 Turn off the battery switch, perform the following setting, and turn on the battery switch again if “E09” is displayed on the temperature display.
SPECIFICATIONS Name Model Pharmaceutical Refrigerator MPR-1411 MPR-1411R External dimensions W1440 x D830 x H1950 (mm) Internal dimensions W1320 x D710 x H1500 (mm) Effective capacity 1364 L 1359 L Exterior Painted steel Interior Painted steel Door Double layer pair glass/steel plate, automatic closing mechanism, 2 doors Insulation Shelf Rigid polyurethane foamed-in place Hard steel wire on polyethylene coating 8 pcs. ----- Painted steel 10 pcs.
CAUTION Please fill in this form before servicing. Hand over this form to the service engineer to keep for his and your safety. Safety check sheet 1. Refrigerator contents : Risk of infection: Risk of toxicity: Risk from radioactive sources: غYes غNo غYes غYes غYes غNo غNo غNo (List all potentially hazardous materials that have been stored in this unit.) Notes : 2.
SANYO Electric Co.