Balance Operating Menu
The balance operating menu enables you to adapt the balance to your requirements.
At the factory, we have set the codes for the balance operating program so that you normally will not need to
make any changes. lf you have special operating conditions, change the codes in the balance operating
program to meet your individual requirements.
The
factory settings
in the balance operating program are marked with "*". Functions not marked by "*" can
be set by entering the corresponding code (see page 26).
lf you need to change the factory settings, make sure to record the date.
Weighing Mode:
Adaptation to Ambient Conditions
The balance is adapted to ambient conditions by changing the time allowed for weighings (see the Product
Data Sheet).
Code Changes
Very stable environment (shortest measuring time) 1 1 1 *
Stable environment (short measuring time) 1 1 2 *
Unstable environment (long measuring time) 1 1 3
Very unstable environment (longest measuring time) 1 2 4
Stability Range
The stability symbol remains displayed for a weight
deviation ±
Code Changes
0,25 digit 1 3 1
0,5 digit 1 3 2
1 digit 1 3 3 *
2 digits 1 3 4 *
4 digits 1 3 5
8 digits 1 3 6
Tare Parameters
The tare function is performed:
Code Changes
Without stability control 1 5 1
At stability 1 5 2 *
Auto Zero Function
When this function is activated, any zero point is eliminated so that a true zero indication is maintained.
Code Changes
Auto-Zero on 1 6 1 *
Auto-Zero off 1 6 2
* = factory setting, depends on balance model in some cases