Operating instructions

35
Specifications
Model BA 210 S BA 160 P BA 110 S BA 61
Weighing range structure SuperRange PolyRange SuperRange
Weighing range/capacity g 210 30/60/110/160 110 61
Readability
mg 0,0001 0,0001/0,0002/
0,0005/0,001
0,0001 0,0001
Tare range (by subtraction) g -210 -160 -110 -61
Reproducibility
(standard deviation)
g
±0,0001
±0,0001/0,0002/
0,0005/0,001
±0,0001
±0,0001
Linearity g
±0,0002
±0,0002/0,0002/
0,0005/0,001
±0,0002
±0,0002
Response time (typical) g
2,5
2
2
2
Adaption to ambient condi-
tions and application require-
ments
←
by selection of 1 of 4 optimized filter levels
→
Display update (depends on
the filter level selected
s
←
0,1 – 0,4 (selectable)
→
Allowable ambient tempera-
ture range
o
C
←
+5 .. +40
→
Operating temperature range
o
C
←
+10 ... +30
→
Sensitivity drift within
+10 ... +30
o
C
/C
o
±2 10
-6
±3 10
-6
±2 10
-6
±2 10
-6
Moisture-proof rating
according to DIN 40 040
←
Class F, non-condensing
→
Pan size mm
←
Ø 80
→
Wägeraumhöhe mm
←
200
→
Net weight, approx. kg
←
5,5
→
Selectable weight units g, kg, ct, lb, oz, ozt, tlh, tls, tlt, GN, dwt, mg, /lb, tlc, mom, K, tol, bat, MS
Automatic zero tracking
function (can be turned off by
menu code)
←
standard feature
→
Standard calibration weight
(min. accuracy class)
g 200 (E2) 100 (E2) 100 (E2) 50 (E2)
Other permissible calibration
weight
(min. accuracy class)
g 100,
100+50 (E2)
50,
100+50 (E2)
50 (E2) - - -
AC power source/
power requirements
V~
←
via AC adapter, 230 or 115, -20% ... +15%
→
Frequency Hz
←
48 – 60
→
Power consumption, (typical) VA
←
maximal 16; typical 8
→
Hours of operation with a fully-
charged YRB 02 Z external
battery pack
h
←
48
→
Built-in interface
RS 232 C-S/V24-V28; 7-bit; -even, -odd;
transmission rates 150 … 19200 Baud,
1 or 2 stopbits, software/Hardware handshake