Example
Calibrate/adjust the scale
Settings: factory settings
Step Press key Display/Printout
1. Turn on scale, if power is off [ON/OFF]
0.0 g
2. If necessary, zero the scale [ZERO] 0.0 g
3. Start calibration/adjustment [TARE] (> 2 sec) + 500.0 g
Calibration weight is
displayed without wt. unit
4. Place the indicated calibration
500.0 g
weight on scale (in this case,
500 g)
After calibration/adjustment,
+ 500.0 g
calibration weight is displayed
with wt. unit
5. Remove calibration weight