Operating instructions
31
Data Interface
Purpose
Your CB scale comes equipped with an interface
port for connection to a computer or other peripheral
device.
You can use an on-line computer to change, start
and/or monitor the functions of the scale and the
application program (such as counting).
Features
Type of interface: serial interface port
Operating mode: full duplex
Standard: RS-232
Transmission rates: 150; 300; 600; 1,200;
2,400; 4,800; 9,600 baud
Parity: mark, space, odd, even
Character format: 1 start bit, 7-bit ASCII,
parity, 1 or 2 stop bits
Handshake: 2-wire interface:
via software (XON/XOFF)
4-wire interface:
via hardware (CTS/DTR)
Communications mode: SBI
Data output format: 16 or 22 characters