Operating instructions
33
Data Output Format with 16 Characters
Display segments that are not activated are output as spaces. Characters without a decimal
point are output without a decimal point.
The following characters can be output, depending on the characters displayed:
Normal Operation
Position 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
+ DDD DDD * UUUCRLF
or – ..... ... ***
or * ***** ***
or 0 0 0 0 0 0
*: Space CR: Carriage return
D: Character displayed (digit or letter) LF: Line feed
U: Unit symbol
Special Codes
Position 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
**––******CRLF
or H *
or L *
or C *
*: Space H: Overload
– –: Unstable weight L: Underload
(final readout mode: C: Calibration/Adjustment
only the stable weight is shown in digits)
Error Codes
Position 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
**U*###****CRLF
*: Space
# # #: Error code number