Operating instructions
34
Example: Output weight + 1255.7 g
Position 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
+** *125 5.7*g**CRLF
Position 1: Plus or minus sign or space
Position 2: Space
Positions 3–10: Weight with a decimal point; leading zeros are output as spaces
Position 11: Space
Positions 12–14: Unit symbol or space
Position 15: Carriage return
Position 16: Line feed
Data Output Format with 22 Characters
When data with an ID code is output, the ID code consisting of 6 characters precedes the
data with the 16-character format.  These 6 characters identify the following value:
1234567891011121314151617181920 21 22
IIIIII+*DDDDDDDD * UUU  CR  LF
*****– ........ ***
* ********
000000
I: ID code character
1)
U: Unit symbol
*: Space CR: Carriage return
D: Character displayed LF: Line feed
1)
depends on scale type