Operating instructions
38
Software Handshake
The software handshake is controlled via XON and
XOFF.  When a device is switched on, XON must 
be transmitted to enable any connected device to 
communicate.
Activating Data Output
You can define the data output parameter so that 
output is activated either when a print command is
received or automatically and synchronous with the
scale display or at defined intervals (see application
program descriptions [“Counting”] and automatic 
print setting).
Data Output by Print Command
The print command can be transmitted by pressing
[PRINT] or by a software command (Esc P).
Automatic Data Output
In the “automatic print” mode, data is output to 
the interface port without a print command.  You can
choose to have data output automatically at defined
print intervals with or without the stability parameter.
Whichever parameter you select, the data will be output
as the readouts appear on the scale display. The display
update frequency depends on the menu setting for
“Adapt filter” 1.1.x.
If you select the automatic print setting, data will 
be transmitted immediately the moment you turn 
on the scale.