Operating instructions
41
Error Codes
Error codes are shown on the main display for approx. 2 seconds, after which the program
automatically returns to the weighing mode.
Display Cause Solution
No segments appear No AC power is Check the AC
on the display available power supply
The AC adapter  Plug the AC adapter
is not plugged in into an electrical outlet
Automatic shutoff Turn on the scale
is configured
Non-rechargeable/ Replace or recharge
rechargeable batteries using external
batteries are drained charger
h The load exceeds  Unload the weighing pan
the scale capacity 
l Something is touching  Move the object
the weighing pan that is touching the 
weighing pan away
e 01 Data output is not Change the setting
compatible with to the compatible one
the input format in the menu
e 02 Calibration/adjustment Calibrate/adjust only 
condition not met, e.g.: when zero is displayed
– scale not zeroed Press [ZERO] to 
zero scale display
– scale is loaded Unload scale
e 08 The scale was zeroed Only zero the scale 
outside the zero range when in the allowable
zero range of ± 2% of 
the maximum capacity
e 09 Taring not possible Press [ZERO] 
when the gross value to zero the scale
< zero