Display Cause Solution
e 22 The weight is too light Increase the reference
or there is no sample  sample quantity
on the weighing pan
e 30 The interface port  Set code to reset menu or
for printer output contact your local
is blocked Sartorius Service Center
The weight readout Unstable ambient Set up the scale in 
changes constantly conditions (too much a different area or
vibration or the scale) change the menu setting
is exposed to a draft) to adapt the scale
A foreign object is Remove the foreign 
caught between the object
weighing pan and 
the scale frame
The weight readout The scale was not  Calibrate/adjust the scale;
is obviously wrong calibrated/adjusted;  zero before weighing
the scale was not
zeroed before weighing
If any other errors occur, please contact your local Sartorius service center:
for a list of the addresses, please visit our Internet website at: http://www.sartorius.com 