Operating instructions
Connecting Electronic Peripheral Devices
§ Make absolutely sure to unplug the balance/
scale from AC power before you connect or
disconnect a peripheral device (printer or PC)
to or from the interface port.
Warmup Time
To deliver exact results, the balance/scale
must warm up as listed below after initial
connection to AC power or after a relatively
long power outage.
– Model CP2P microbalance: at least 4 hours
– All other precision and analytical models:
at least 30 minutes
Only after this time will the balance/scale
have reached the required operating temper-
ature.
Using Verified Balances/Scales as Legal
Measuring Instruments in the EU*
$ The balance/scale must warm up for at
least 24 hours after initial connection to
AC power.
Leveling the Balance/Scale
Purpose
– To compensate for unevenness at the place
of installation.
– To achieve perfectly horizontal positioning
of the balance/scale for consistent
reproducibility
Always level the balance/scale again after
any time it has been moved.
* including the Signatories of the Agreement
on the European Economic Area
13