Operating instructions
18
Below-Balance/Under-Scale Weighing
A port for a below-balance/under-scale
weighing hanger is located on the bottom of
the balance/scale. You can order the hanger
directly from Sartorius for balances/scales
with a weighing capacity over 12 kg
(see “Accessories").
● Open cover plate on the bottom
of the balance/scale.
● Using the built-in hanger (1): Attach the
sample (e.g., using a suspension wire) to
the hanger
$ Bore hole (2): Carefully fasten the special
hanger, or order a hanger directly from
Sartorius.
$ If necessary, install a shield for protection
against drafts
Important Note Concerning Verified
Balances/Scales Approved for Use as Legal
Measuring Instruments in the EU*:
The below-balance/under-scale weighing
port may not be opened or used when an
approved balance/scale is being operated
as a legal measuring instrument.
● Remove the brass screw
● Suspend the below-balance weighing hook
supplied from the hanger
● If necessary, install a shield for protection
against drafts
* including the Signatories of the Agreement
on the European Economic Area
2
1
Analytical and Precision Balances:
Microbalance CP2P..: