Operating instructions
23
External Calibration
Settings:
Calibration/adjustment mode: External calibration/adjustment (menu code 1 9 1)
The weight required for calibration/adjustment is defined in the factory settings
(see “Specifications").
Step Key (or instruction) Display/Printout
1. Turn on the balance/scale, e
0.0 g
if necessary.
2. Tare the balance/scale, t
0.0 g
if necessary.
3. Start calibration/adjustment. q
+5000.0 g
CAL G
4. Apply the prompted calibration
weight (in this example:
5000 g).
5. Balance/scale is calibrated
– 0.2 g
(displayed only if menu CAL ± G
code 1 10 2 is set).
6. If the menu code for
Adjust*
“calibration and adjustment CAL
in a single step" (1 10 1)
is set, the balance/scale is
now adjusted automatically.
7. At this point, adjustment
CC
is completed. CAL
8. After calibration, the weight
+ 5000.0 g
unit is shown.
9. Remove the calibration weight.
0.0 g
* displayed only if menu item 1 10 2 is selected.