Operating instructions

31
Setting IDs, Time, Date
and Display Brightness
Purpose
Configuring measurement environment
parameters for ISO/GLP-compliant
data records.
Setting date and time (for ISO/GLP-
compliant records only). Adapting display
to ambient lighting conditions.
Features
Enter up to 8 characters to identify a
measurement series. Permissible characters
include the numbers 0 through 9, the dash
or minus sign (“–") and spaces. Leading
zeroes are output as spaces.
Date and time of beginning and end
of series recorded.
Display brightness
1
):
0 = off, levels of brightness: 1 through 9
Activate the ID number, date and time
configuration menu:
Press e to turn the balance/scale
off and back on again. While all display
segments are lit, press v briefly
Scroll upward : press q
Scroll right : press p/r
Press t to confirm input and toggle
between ID number, time and date.
Store settings and exit menu:
Press and hold t (2 sec.)
1
) No display backlighting on models CP2P..,
CP225D and CP225D-0CE