Operating instructions

Counting
39
Menu code: 2 1 4
Display symbol: Z
Purpose
With the Counting program you can
determine the number of parts that each
have approximately equal weight.
Features
The minimum load is equal to one digit,
defined according to the resolution of the
active weight unit.
Press and hold the v key (2 seconds)
to set the reference sample quantity.
Configure the resolution used when
reference sample quantity is stored and
piece counts are calculated.
Optional automatic output of the piece
count and average piece weight to the
data interface port when the menu
code 7 1 2 (print application parameters)
is set.
Long-term storage of the last reference
sample quantity “nRef" entered.
Toggling between piece count and
weight by pressing v.
Function Keys
v:
Begin determination of piece weight
> Application program initialized with
predefined reference sample quantity.
c:
End application program; clear
initialization data
Changing the reference sample quantity:
§ Press and hold v (2 sec.)
> Current reference sample quantity
is displayed.
$ Press v briefly to change the value;
press repeatedly until the desired reference
sample quantity is displayed.
Quantities to choose from: 1, 2, 5, 10, 20,
50, 100.
§ Store setting in long-term memory:
Press and hold v.