Operating instructions

40
Reference Sample Updating
Automatic reference sample updating
optimizes the counting accuracy.
You can activate or de-activate this
function in the menu.
Automatic reference sample updating is
performed when:
the criterion for the stability parameter
selected in the menu has been met
the current piece count is less than twice
the original piece count
the current piece count is less than
100 + the initialization value
the internally calculated piece count
(such as 17.24 pcs) differs by less than
± 0.3 pcs from the whole number
(17 pcs in this example)
> The abbreviation opt, for “optimizing",
is displayed briefly with the new reference
sample quantity.
Reference Weighing
(Counting with two balances/scales)
Purpose:
Use of a reference balance/scale affords
higher precision in counting large amounts
of parts. The CP balance is used to deter-
mine the reference weight. The following
Sartorius terminals can be used for sample
weighing in conjunction with a reference
balance/scale:
For QC scales menu code 8 8 2
For FB/FC/LA/LP series balances/scales,
menu code 8 8 3
For isi terminals menu code 8 8 4
§ Please order the required connecting cables
directly from Sartorius
The following settings must have the same
configurations in both balances/scales:
“Counting” program
Weight units
Settings in the CP balance:
Set menu codes 7 1 2 and 7 2 2
All data interface parameters
Baud rate
Parity
Number of stop bits
Handshake mode
Transfer Function in the CP Balance:
Press the v key
> The reference value is transferred to the
counting balance/scale
Counting Balance/Scale:
$ See the operating instructions of the
balance/scale in question for details