Example: Determining residual weight in percent
Settings:
Menu: Weighing in percent program (menu code 2 1 5),
Print application parameters: On; all parameters (menu code 7 1 2),
Line format for printout: 22 characters (menu code 7 2 2)
Reference percentage: ref 100%
Step Key (or instruction) Display/Printout
1. Turn on the balance/scale, e
if necessary.
2. Place empty container
+ 22.6 g
on balance/scale.
3. Tare the balance/scale. t
0.0 g
4. Place sample equal to 100%
of reference percentage
on the balance/scale (in this
example: 111.6 g).
5. Initialize the balance/scale. v
ref 100 (displayed briefly)
+ 111.6 g
+ 100.0 %
pRef + 100 %
Wxx% + 111.6 g
6. Remove container; for example,
to treat sample (in this example,
the sample is now dried).
7. Place container with sample
+ 94.9 %
on the balance/scale again
(after treatment).
8. Optional: print percentage. p/r
Prc + 94.9 %
9. Display residual weight and c + 105.9 g
delete reference value.
10. Optional: print net p/r
N + 105.9 g