Operating instructions

56
Output Format with 16 Characters
Display segments that are not activated are output as spaces.
The following characters can be output, depending on the characters displayed on
the balance/scale:
Normal Operation
Position 1 2 3 45678910111213141516
+ DDDDDD * UUUCRLF
or ........ ***
or * ********
*: Space CR: Carriage return
D: Digit or letter LF: Line feed
U: Unit symbol
Special Codes
Position 1 2 3 45678910111213141516
**********CRLF
or H *
or L *
or C *
*: Space H: Overload
C: Calibration/adjustment L: Underload
Error Codes
Position 1 2 3 45678910111213141516
Err*###****CRLF
*: Space
# # #: Error code number