Operating instructions

Pin Assignment Chart
62
Female interface connector:
25-contact D-Submini DB25S with screw-lock hardware
Male connector used (please use connectors with the same specifications):
25-pin D-Submini DB25 with integrated shielded cable clamp assembly
(Amp 826 985-1C) and fastening screws (Amp type 164868-1)
! Warning When Using Pre-wired RS-232 Connecting Cables:
RS-232 cables purchased from other manufacturers often have incorrect pin assignments
for use with Sartorius balances/scales. Be sure to check the pin assignment against
the chart below before connecting the cable, and disconnect any lines marked “Internally
Connected” (e.g., pin 6). Failure to do so may damage or even completely ruin your
balance/scale and/or peripheral device.
Pin Assignment Chart:
Pin 1: Shield
Pin 2: Data output (TxD)
Pin 3: Data input (RxD)
Pin 4: Internal ground (GND)
Pin 5: Clear to send (CTS)
Pin 6: Internally connected
Pin 7: Internal ground (GND)
Pin 8: Internal ground (GND)
Pin 9: Reset _ In *)
Pin 10: Not connected
Pin 11: +12 V
Pin 12: Reset _ Out *) For remote switch
Pin 13: + 5 V
Pin 14: Internal ground (GND)
Pin 15: Universal remote switch
Pin 16: Not connected
Pin 17: Not connected
Pin 18: Not connected
Pin 19: Not connected
Pin 20: Data terminal ready (DTR)
Pin 21: Ground input for external supply voltage
Pin 22: Not connected
Pin 23: Not connected
Pin 24: Ext. supply voltage input + 12 ... 30 V
Pin 25: + 5 V
*) = Hardware restart