Display Cause Solution
err 235 Connecting cable not Connect the cable correctly
on the CP225D connected correctly
Connection to junction Connect the equipment
on a different balance/scale correctly
The weight readout Unstable ambient Set up the balance/scale
changes constantly conditions in another area
A foreign object is caught Remove the foreign object
between the load plate and the
balance/scale frame
The weight readout The balance/scale has not Calibrate/adjust the
is obviously wrong been calibrated/adjusted balance/scale
The balance/scale was not Tare or zero the balance/scale
zeroed before weighing before weighing
If any other errors occur, contact your local Sartorius Service Center!
Contact information: Please point your Internet browser to: www.sartorius.com
Preparing CP2P.. Models for Transport
Model CP2P:
§ Remove the weighing pan
§ Remove the interior draft shield from the chamber:
Use your fingers to raise the draft shield carefully
until it is free
$ Place these parts in the accessory kit
§ Fasten the chamber doors by placing a rubber band
around the door handles.
Model CP2P-F:
§ Remove the filter pan lid from the chamber
§ Use forceps to remove the filter weighing pan care-
fully from the chamber
§ Remove the interior draft shield from the chamber:
Use your fingers to raise the draft shield carefully
until it is free
$ Place these parts either in the accessory kit or in the
original packaging
The balance/scale must be conditioned again any
time it is set up in a new location (see “Installation
Instructions” on page 6).