Operating instructions

67
Cleaning the Weighing Chamber and
Draft Shield
§ Open the draft shield cover and take out the
removable parts
§ Use a hand-held vacuum cleaner and
mini-hose to remove any powdered sample
material carefully.
§ Use blotting paper to remove any liquid
sample material.
§ On models with a 3-sided draft shield, pull
the 3 draft shield walls upwards to remove,
if necessary.
Safety Inspection
If there is any indication that safe operation
of the balance/scale with the AC adapter
is no longer warranted:
§ Turn off the power and disconnect the
equipment from AC power immediately
> Lock the equipment in a secure place to ensure
that it cannot be used for the time being
Safe operation of the balance/scale with the
AC adapter is no longer ensured when:
there is visible damage to the AC adapter.
the AC adapter no longer functions properly.
the AC adapter has been stored for a relatively
long period under unfavorable conditions.
Maintenance and repair work may be
performed only by service technicians who are
authorized by Sartorius and who
have access to the required maintenance
manuals.
have attended the relevant service training
courses.
We recommend having the power supply
inspected by a certified electrician at regular
intervals, according to the checklist given
below:
Insulating resistance > 7 megaohms measured
with a constant voltage of at least 500 V at
a 500 kohm load
Leakage current: < 0.05mA measured with
a properly calibrated multimeter