Operating instructions
Specifications
Verified models with EC type approval
Competence Series/CP Series
Model CP225D-0CE CP324S-0CE CP224S-0CE, CP124S-0CE, CP64-0CE
CP224S-PCE CP124S-ACE
Type BC BL 100 BC BL 100 BC BL 100 BC BL 100 BC BL 100
Accuracy class
1
) k kkkk
Maximum capacity, Max.
1
) g 80/220 320 220 120 64
Scale interval, d
1
) mg 0.01/0.1 0.1 0.1 0.1 0.1
Tare range (subtractive) g < 100% of the maximum capacity
Verification scale interval, e
1
) g 0.001 0.001 0.001 0.001 0.001
Minimum capacity, Min.
1
) g 0.001 0.01 0.01 0.01 0.01
Stabilization time (average) s < 6/3 < 3 < 2 < 2 < 2
Range of use g 0.001–220 0.01–320 0.01–220 0.01–120 0.01–64
(according to CD
1
)
Allowable ambient
operating temperature °C +15… +25 (59°F to 77°F)
External calibration weight g 200 (E2) 200 + 200 (E2) 100 (E2) 50 (E2)
(of at least accuracy class) 100 (E2)
Net weight, approx. kg 7.6 6.5 6.5 6.5 6.5
Pan size (inner diameter)* mm 80 d
Pan area* cm
2
64
Weighing chamber height
(from pan to sliding cover) mm 232
Dimensions (W+D+H) mm 213+342+340
Electronics box (W+D+H) mm 134+51+155––––
AC power source/ V~ AC adapter STNG6 230 VAC or 115 VAC,
Power requirements +15%…–20% (protection rating IP20)
Frequency Hz 48 – 60
Power consumption (average) VA maximum: 16; average: 8
Approx. hours of operation
with the YRB05Z
rechargeable battery pack h 20 22 22 22 22
Selectable weight units Grams, milligrams, carats Grams, carats
Built-in interface RS-232C-S/V24-V28; 7-bit;
parity: even, mark, odd, space;
transmission rates 150...19,200 baud;
1 or 2 stop bits; software/hardware handshake
1
) CD = Council Directive 90/384/EEC on non-automatic weighing instruments;
applicable to the European Economic Area
* Three-sided weighing pan: d = Diameter of inner circle.
The cross-hatched section can be fully utilized.
76