Specifications

10
Cubis display and service units
Select the display and service unit and enter it in the field marked with the icon.
Types MSA MSU MSE
Cubis leveling
Select the type of leveling and enter the identifier “Ø” or “1” in the field marked by the icon.
Ø The Cubis shows the level indicator on the display and provides support for rapid leveling
(as standard on the display and service units MSA and MSU; only a warning message in MSE).
1 Fully automatic, motorized Q-Level leveling at the touch of a button (available for all analytical
balances and semimicrobalances with 0.1 mg or 0.01 mg readability and all precision balances with
1 mg readability).
Test certificates and permits
Select a test certificate | permit and enter the identifier in the field marked with the icon.
ØØ Standard certificate of conformity to specifications
TR Like ØØ, but with a detailed test protocol
CE Calibrated in the factory with European calibration permit
Optional interface modules
Depending on the balance, it may be possible to select an additional interface module.
IR RS232 interface, 25-pin
IB Bluetooth
®
interface
IP RS232 interface, 9-pin, inc. PS/2 interface
Please use the adjacent fields to enter the selection
made with the icon.
Operation
Display
Adaptation of the
service unit
Standard data interfaces
SD card reader
Operation of the
motorized draft shield
(only for DA or DI
draft shields)
Applications
Touchscreen, keys for central basic
functions
High-resolution color TFT,
5.7” graphical display
Tiltable display,
removable service unit
– USB (integrated into weighing module)
– RS232C accessory interface,
25-pin (integrated into weighing module)
– Ethernet (integrated into display and
service unit)
Integrated as standard into display and
service unit
Actuation via side keys or contactlessly
via IR switch (optional), learning
capability
Unit conversion, SQmin function for
minimum initial weight according to
USP, isoCAL automatic calibration |
adjustment function, individual identi-
fiers, density determination, statistics,
calculations, averaging, formulation,
weighing in percent, time-controlled
functions, totalizing, DKD measurement
uncertainty, second tare memory, count-
ing, checkweighing
Keys
High-resolution black-and-white,
5.7” graphical display
Tiltable display,
removable service unit
– USB (integrated into weighing module)
– RS232C accessory interface,
25-pin (integrated into weighing module)
– Ethernet (integrated into display and
service unit)
Integrated as standard into display and
service unit
Actuation via side keys or contactlessly
via IR switch (optional), learning
capability
Unit conversion, SQmin function for
minimum initial weight according to
USP, isoCAL automatic calibration |
adjustment function, individual identi-
fiers, density determination, statistics,
calculations, averaging, formulation,
weighing in percent, time-controlled
functions, totalizing, DKD measurement
uncertainty, second tare memory, count-
ing, checkweighing
Keys
Black-and-white LCD display,
removable service unit
– USB (integrated into weighing
module)
– RS232C accessory interface,
25-pin (integrated into weighing
module)
Actuation via keys or contactlessly
via IR switch (optional), learning
capability
Unit conversion, SQmin function for
minimum initial weight according to
USP, isoCAL automatic calibration |
adjustment function, density determi-
nation (buoyancy method only), calcu-
lations, averaging, net | total formula-