Specifications

Premium Semimicrobalances and Analytical Balances:
Sartorius ME – When Results Matter
13
Unrivaled speed
An outstanding feature of the Sartorius ME
series is speed: stable readouts with five
decimal places in just eight seconds.
Controlled by palm-operable keys or by cus-
tom programming, the draft shield closes
quietly, precisely and quickly. It can be
adapted to any weighing scenario, accord-
ing to the requirements of your application.
Unrivaled stability
Repeatability of the weights measured is an
additional strength of the Sartorius ME.
The results are just as stable as the mono-
lithic weighing system of the ME.
The Sartorius ME is amazingly impervious
to the surrounding environment of its
installation location. Interfering static elec-
tricity on samples and tare containers can
be neutralized at the touch of a button.
Unrivaled reliability
Sartorius ME stands for reliability, year in,
year out. That’s why we are offering a three-
year warranty, which we will extend on
request up to a total of five years.
The Facts
Unique monolithic weighing system
Three-part, motorized draft shield system
User-friendly palm-activated keys for draft
shield operation; foot switch optional for
applications where you need your hands free
Neutralization of static electricity
Prompts in clear English for operator
guidance
Alphanumeric input capability for sample
IDs
Software support for use in quality
management systems
SQmin function displays the minimum ini-
tial weight in accordance with the United
States Pharmacopeia (can be activated by
Sartorius Service)
Display of measurement uncertainty
according to the German Calibration
Service (DKD)
ISO|GLP-compliant, user-configurable
records|printouts
Built-in applications
Built-in software supports all key laboratory
weighing applications to ensure smooth,
time-saving lab procedures and reliable
results.
– Density determination
– Calculation of weights using a definable
factor or equation
– Statistical evaluation
– Differential weighing
– Air buoyancy correction
– Air density determination
Specifications
Model ME235S ME235P ME235P-SD* ME614S ME414S ME254S
Weighing capacity (g)230 60|110|230 60|110|230 610 410 250
Readability (mg) 0.01 0.01|0.02|0.05 0.01|0.02|0.05 0.1 0.1 0.1
Repeatability 0.015 0.015 0.015 0.1 0.1 0.07
(<mg) (0–60 g) (0–60 g) (0–60 g)
0.025 0.040 0.040
(60–230 g) (60–110 g) (60–110 g)
0.040 0.040
(110–230 g) (110–230 g)
Linearity (< ± mg) 0.1 0.15 0.15 0.4 0.3 0.15
Response time (<s) 8 8 8 3 2.5 2.5
Off-center load at 1 0.15 0.2 0.2 0.6 0.4 0.3
maximum load (<mg)
(according to OIML R76)
Weighing pan 90 90 90 90 90 90
diameter (mm)
Clearance above 253 253 195 253 253 253
weighing pan (mm)
All models are available in verified versions (excluding *)
* with short-design draft shield and pipette opening, d 60 mm, with cover