Specifications

53
Specifications|Accessories
LMA300P
Specifications*
Measuring range (%) Approx. 0.1–60
Readability (%) 0.01
Measuring accuracy (%) ± 0.05
(depends on calibration and type of sample)
Measuring time (s) < 1
Display % moisture, % dry weight (solids)
Measurement method Microwave resonance technology
Allowable sample temperature (°C) Approx. 0–70
Operator guidance Touch screen with demand-driven menu based on alphanumeric prompts
(dialogue text and symbols)
Memory for number of
measurement programs 40
Data printer, optional External
GLP-compliant report Yes, with optionally available printer, YDP20-0CE
Interface port 2 + RS-232 C for printer and PC
USB port + 128 MB USB flash drive
Line voltage 110–230 V AC
Frequency 50 ... 60 Hz
Power consumption 60 VA max.
Housing dimensions (in mm) W + D + H
Control unit, LMA300PA 500+430+200
Weight approx. (kg)
Control unit, LMA300PA 11.5
* * In addition to the LMA300PR sensor module, other sensors are also available on request. Depending on the desired application,
however, the technical specifications will have to be agreed on with a Sartorius applications technician.
Optional Accessories Order no.
Data printer YDP20-0CE
Ink ribbon cassette for YDP03-0CE 6906918
Printer paper, 5 rolls, each with 50 m,
for YDP03-0CE 690693
Applicator, 60 mm 69MA0294
Applicator, 140 mm 69MA0295
Reference standard LMA301SY
Sensor Specifications LMA300PR LMA301PR LMA302PR LMA303PR LMA304PR
Dimensions (mm) 370+245+275 370+245+275 370+395+375 260+270+280 370+385+375
Weight 10 kg 10 kg 11 kg 5 kg 15 kg
Sample volume (60/150) ml (90/125) ml 400 ml 27 ml 2000 ml
Resonator diameter 40 mm 50 mm 46 mm 26 mm 96 mm