Example
Determination of the Air Density by Weighing, Using the Reference Weights Supplied for Steel and Aluminum
Settings (changes in the factory settings required for this example):
Setup: Application parameters: Application 1: Air buoyancy correction: Air buoyancy determination: Yes
Air buoyancy correction: Change steel/aluminum references: Yes
Step Press key(s) (or follow instructions) Display/Output
1. Turn on the balance if it is e
not already on
2. Start air buoyancy correction
Start soft key
3. Toggle to the mode for air density
RhoA soft key
determination to enter the
specifications for steel and aluminum
4. Toggle to the display of the
Param. soft key
air density parameters
5. Toggle to the display of the
Ref. soft key
air density references
6. Enter the specification of the steel 200.
reference supplied (in this case, 00821
200.00821 g/cm
3
)
7. Confirm entry
l soft key
8. Enter the specification of the 200.
aluminum reference supplied 00348
(in this case, 200.00348 g/cm
3
)