Operating instructions

Diameter Determination
102
Purpose
This application program enables you to
determine the diameter of round wires
and metallic filaments. This program
is used, for example, to determine the
diameter of filaments for light bulbs.
You can use density determination in
combination with a program chosen
from Application 2 (checkweighing,
time-
controlled functions) and Application 3
(totalizing, formulation, statistics).
Features
Input and changing of individual data
after pressing
Param. soft key
Identifier 1
Text1
(20 characters max.)
Identifier 2
Text2
(20 characters max.)
Density of the sample,
RhoWt
(0.01 to 50.,0 g/cm
3
;
Factory setting = 8.0 g/cm
3
)
Length of the sample in millimeters
(range = 0.1 to 99999 mm)
Number of decimal places for the dia-
meter result (0 7; factory setting = 3)
Input of the density and length of the
sample directly using the numeric keys
When this program is active, the
following information is displayed
in the text line:
Data record name (if you have selected
the “Product data memory” function)
Density of the sample
RhoWt
Length in millimeters mm
The calculated diameter result is indi-
cated by the “mm” ID
Automatic initialization of this applica-
tion when the power is turned on and
loading of the stored data record, pro-
vided that all data were entered, and
automatic power-on initialization, pro-
vided this has been selected in the Setup
menu (Setup: Application parameters:
Auto-start application when power goes
on: Yes)
The start page is skipped and the diam-
eter determination program is then
immediately activated.
Extra function “Product data memory”:
up to 300 data records can be saved for
the diameter determination program
(for additional functions, please see
page 123)
Press c to end diameter
determination
Soft Key Functions
Start Starts diameter determination
Param. Begins input of diameter and
length
RhoWt Saves the density of the
sample
l(mm) Saves the length of the sample
Weigh. Display the weight
Dia. Displays the calculated
diameter result
Preparation
§ Turn on the balance: press e
> The Sartorius logo is displayed
§ Configure the “Density detemination”
program in the Setup menu:
press o
§ Select the
Application
parameters:
press the
q soft key 2 times, then the
O soft key
§ Select the
Application
parameters(basic set-
tings): press O soft key
§ Select
Diameter determi-
nation: press the Q or q soft key,
repeatedly, if necessary
§ Confirm
Diameter determi-
nation: press the O soft key
§ Save settings and exit the Setup menu:
press the
oo soft key
Additional Functions
In addition to the functions for:
alphanumeric input
taring (not during alphanumeric input)
printing,
you can also access the following
functions from this application:
Calibration/Adjustment
§ Press the
CAL soft key
> See “Calibration/Adjustment”
for further instructions
Toggling to the Next Application
§ Press d
> See the section on the corresponding
application program for further
instructions
Setup (Setting Parameters)
§ Press o
> See “Configuring the Balance”
for further instructions
Turning Off the Balance
§ Press e
> The balance shuts off