Operating instructions

31
Device Parameters (Overview)
ο factory setting
user-defined setting(s)
Setup –
Device
parameters
Draft shield Keys ο Same function
Left/right key Separate function
Off
Automatic mode ο Off
Close –>
function –> open
Close –>
perform function
Weight ο Show all decimal places
resolution Do not show last decimal place
with door open
Ionizer* Off
ο On Auto-off time: 10 s 1 .. 99 seconds
Password:
None
Enter 8 characters
max.
User ID:
None
Enter 20 characters
max.
Clock Time: Enter
hh.mm.ss
Date: Enter dd.mm.yy
or mm.dd.yy (U.S. mode only)
(01.01.97)
Interfaces Serial ο SBI Baud rate 150 baud
communication 300 baud
(PERIPHERALS) 600 baud
ο 1200 baud
2400 baud
4800 baud
9600 baud
19,200 baud
Number of data bits ο 7 bit
1)
8 bit
Parity Space
2)
o Odd
Even
None
3)
Number of stop bits ο 1 stop bit
2 stop bits
Handshake mode Software handshake
ο Hardware handshake,
1 character after CTS
xBPI Network address: ο Enter any number from 0 to 31
For the display, keys and extra functions, see next pages
* = not on ME36S, ME5, SE2
1)
not if “None” parity is selected
2)
only if 7 data bits selected
3)
only if 8 data bits selected
Factory setting
Factory setting
Factory setting