Operating instructions

63
isoCAL:
Automatic Calibration, Adjustment
and Linearization
First set either
On and reset the
application, isoCAL
and linearization on
or On without resetting
the app. (factory setting) in
the “Setup: Balance/scale functions”
menu.
The “isoCAL” display automatically
begins flashing if the ambient
temperature changes in relation to the
temperature at the time of the last
calibration/adjustment, or after a
defined time interval has elapsed.
The balance is telling you that it wants
to adjust itself.
This automatic calibration and adjust-
ment prompt is activated when:
The change in temperature is greater
than1.5 Kelvin or the elapsed time
interval is greater than 4 hours
The balance Setup mode is not active
No number or letter input is active
The load has not been changed within
the last 2 minutes
The balance has not been operated
within the last 2 minutes
The load on the balance does not
exceed 2% of the maximum capacity.
When you turn on the balance after
it has been disconnected from power
(only on verified balance)
When these requirements are met,
C is displayed in the line for measured
values.
If the balance is not operated and
the load is not changed, internal
calibration and adjustment will start
after 15 seconds have elapsed.
Automatic Calibration and Adjustment
at Set Times *
In the Setup menu, you can now enter
up to three different times of day for
automatic calibration/adjustment (see
menu tree on page 55). When one
of these times is reached, the balance
displays the flashing calibration prompt
(“isoCAL”). Calibration/adjustment is
not performed if the balance is
off (standby) or
in the Setup mode at the time set for
calibration.
If the balance is being operated at
the time set for automatic calibration/
adjustment, the calibration/adjustment
sequence will not be prompted or
performed afterwards.
Automatic calibration/adjustment
is prompted at set times when:
The set time is reached
The balance Setup mode is not active
No alphanumeric input is active
(e.g., equation for calculation)
The load has not been changed within
the last 2 minutes
The balance has not been operated
within the last 2 minutes
The load on the balance does not
exceed 2% of the maximum capacity
* = does not apply to verified balances
In the Setup menu, you can configure
the balance so that after calibration and
adjustment
the application program
is restarted
On and reset
the application
isoCAL and linearization are activated
and the application program must be
restarted
isoCAL and
linearization on
the application program remains at its
previous status
On without resetting
the app.
Also in Setup, you can configure the
balance so that it displays a calibration
prompt, but does not perform the
calibration/adjustment functions
automatically:
Only adjustment prompt