Technical data
118
ERWEKA TBH 425
Hardness Tester for automatic measurement of
up to 10 samples. Via the numeric membrane
keypad with alpha-numeric sub function the
nominal hardness in combination with 2 individ-
ual plus/minus tolerances (T2 = pass/fail crite-
ria) and remarks may be stored for up to 20
different products and recalled for testing. In-
formation such as product name and remarks
can be entered similar to writing a message on
a mobile phone. Stored product data can be
transmitted via SD card or USB stick to other
devices of the TBH 325/425 family. Calibration
is completely menu guided and may be per-
formed by either weight stones or the electronic
ERWEKA AutoCal System. Via a standard USB
printer interface a printer can be connected to
receive hardcopy documentation.
The serial LAN/USB interface allows connecting
a computer with the optional MC.net software
for external control. In combination with this, the
TBH 425 series is 100% 21 CFR 11 compliant.
10 chamber star feeder with brushes to
clean the transport way
Routine for alignment of oblong shaped
samples
Tests samples up to 28 mm diameter
Measuring range 10 – 500 N, optional 10 –
1.000 N accuracy 1N
Values in N, Kp or Sc
“Turbo” hardness test jaw, 3 times faster
than previous version
Numeric keypad with alpha-numeric sub-
function for entry of test parameters, prod-
uct name, batch number and remarks
Integrated menu for calibration and ad-
justment with documentation print out
LAN/USB interface for external control and
storage of results by the optional MC.net
software
USB interface for USB stick standard
SD-Card Slot standard
USB printer interface for documentation of
time, date, results and statistics
100V-240V/50Hz and 60Hz 13-445-1000
ERWEKA TBH 425 D
ERWEKA TBH 425 D is based on the TBH 425
but in addition offers measurement and docu-
mentation of the sample diameter/length.
This function allows verifying that the sample
was perfectly aligned before testing hardness.
Additional features:
Measuring range for diameter: up to 28.00
mm
Accuracy 1 N; 0.03 mm
Values in mm or inch
100V-240V/50Hz and 60Hz 13-xx6-xxxx