Technical data
14
ERWEKA DT 720
6, 7 and 8 stirrer type
ERWEKA DT 720 has been designed with the
user in mind: instead of an extensive operation
menu structure and multi-functional keys the DT
720 comes with individual displays for each
test-run parameter (time, rpm, temperature).
UP/DOWN buttons for fast and easy test-run
parameter setting are placed. For better obser-
vation of the sample dissolving process the
one-piece moulded water bath (with standard
water outlet cock) is illuminated (optional).
The manual lifter column, with hydraulic support
for effortless operation, allows raising and low-
ering of the drive head in just a few seconds.
Tablets/Pellets etc. are inserted through the
sampling openings in the drive head via the
standard tablet delivery system. Samples for
analysis are withdrawn through special, mem-
brane sealed sampling openings inside the
drive head. Evaporation less than 0.3% in Low
Head mode and less than 0.7% in High Head
mode (24h/37.0°C).
The extensive use of carbon fibre allowed to
eliminate metal and therefore reduces the risk
of corrosion to a minimum.
The actual test-run parameters can be docu-
mented via an USB printer (optional).
Easy and comfortable operations:
UP/DOWN and START/STOP buttons for
test parameter setting as well as individual
big and bright LE displays for display
Manual Lifter with hydraulic support
Low-Head operations (MagniShaft or Un-
iShaft)
Low-Head- and High-Head operations
(UniShaft only)
Quick exchange system for the stirring
methods by magnets (MagniShaft)
Automatic height adjustment of the stirring
elements
Video monitoring and recording
Automatic vessel centring
One-piece leak-free moulded, illuminated
(optional) water bath with outlet cock
Corrosion-free plate for holding the vessels
Evaporation less than 0.3% in Low Head
mode and less than 0.7% in High Head
mode (24h/37.0°C)
External heating system with control by the
keypad and LE display of the DT 720; ac-
curacy 0.2°C
Standard Tablet Drop Magazine for 8
samples. Optional automatic tablet drop
with control of temperature tolerance
(USP) before drop
Internal PT 100 temperature sensor for
water bath temperature display
Optional external PT 100 temperature sen-
sor for water bath- or media temperature
display
Optional monitoring of the media tempera-
tures by electronic PT 100 temperature
sensors
Calibration menu for quick calibration of
the 1 or 9 temperature sensors
Shaft wobble less than 0.2 mm at delivery
Height adjustable feed for levelling
Automatic self-check of the temperature in
the bath/vessels before test start
Hours of operation counter
RS 232 serial interfaces for external con-
trol
Optional USB printer interface for docu-
mentation of the test run parameters (time,
date, rpm, temperatures etc.) by a stan-
dard printer (PCL 3 language)
Optional USP Sampling Station with auto-
matic height adjustment of the USP sam-
pling points – to create a Dissolution Sys-
tem
Optional video monitoring of six test posi-
tions