Operating Instructions Sartorius Element Series Electronic Precision Balances, Electronic Gem Scale 98648-013-44
Contents 2 Warnings and Safety Precautions 3 Getting Started 7 Operation 7 Basic Weighing Function 7 Description of the Keys 8 8 9 10 11 12 13 Application Programs Toggling between Weight Units Counting Weighing in Percent “Hold” Display Totalizing Specific Gravity 14 Calibration/Span Adjustment 15 Configuration (Setup Menu) 16 Error Codes 17 17 18 19 Overview Specifications Accessories (Options) C Marking Warnings and Safety Precautions Safety Information § To prevent damage to the equipment, please r
Getting Started – – – – Equipment Supplied Balance/scale Weighing pan Plug-in AC adapter Level indicator and adjustable feet Additional equipment with models ELT602, ELT402, ELT202, ELT1001, ELT601: – Calibration weight Additional equipment with model ELT103, GD502: – Calibration weight – Round glass draft shield – Only with GD502: Gem tray – – – – Installation Choose a location that is not subject to the following negative influences: Heat (heater or direct sunlight) Drafts from open windows and doors
! Use only original Sartorius AC adapters: – for Asia: – for the U.S.: 400081.1 400314.1 Connecting the Balance/Scale to AC Power § Insert plug into the jack (located on back of the balance/scale) § Plug the AC adapter into an electrical outlet ! LISTED power supply 11 V–21 V compliant with NEC Class 2 output. Below-Balance/Scale Weighing A port for a below-balance/scale weighing hanger is located on the bottom of the balance/scale. § Open the cover plate on the bottom of the balance/scale.
Using a Non-Rechargeable/Rechargeable Battery (not for model ELT103, GD502) $ A non-rechargeable or rechargeable battery is not included with the equipment supplied. ! Use only a commercially available non-rechargeable (8+ AA/Mignon) or rechargeable battery. ! When using a rechargeable battery, be sure to recharge it outside the balance/scale using a separate charger.
Leveling the Balance/Scale § Always level the balance/scale again any time after it has been moved to a different location. Example: moving bubble from R to L § Turn the feet as shown in the diagram until the air bubble is centered within the circle of the level indicator. > In most cases this will require several adjustment steps.
Operation Basic Weighing Function Features – Zeroing the balance/scale You can zero the balance/scale within the entire weighing range, up to the maximum capacity. Preparation § Switch on the balance/scale: press the e key § If necessary, zero the balance/scale: press the w key $ If necessary, change the configuration settings: see the chapter entitled “Configuration” $ If desired, load the factory settings: see the chapter entitled “Configuration,” menu item 9.
Application Programs Toggling between Weight Units With this application program you can toggle the display of a weight value back and forth between two weight units (see table below). Example: Toggle weight unit from pounds [lb] (application) to grams [g] (basic unit) Step Key (or instruction) 1. Select application program w > 2 sec 2. Select Toggling between Weight Units v 3. Confirm unit 4. Select weight unit; in this example: “5. Pounds” (see table below) 5.
Counting Purpose With the Counting program you can determine the number of parts or items by weighing. Example: Determine the number of uncounted parts; weigh in the selectable reference sample quantity (in this example: 20) Step Key (or instruction) 1. Select application program w > 2 sec 2. Select Counting v repeatedly 3. Confirm setting Symbol “ ” on the display: application is active Display Enter CF 4. Place empty container on the balance/scale 5. Zero the balance/scale w 6.
Weighing in Percent Purpose This application program allows you to obtain weight readouts in percent which are in proportion to a reference weight. Example: Determine an unknown percentage; store the weight on the balance/scale as the reference percentage (100%) Step Key (or instruction) 1. Select application program w > 2 sec 2. Select Weighing in Percent v repeatedly 3. Confirm setting Symbol “ ” on the display: application is active Enter CF 4. Place empty container on the balance/scale 5.
“Hold” Display Purpose The last stable value remains displayed for 5 seconds after removing the sample from the pan. Example: Determine weight of oversized sample Step Key (or instruction) 1. Select application program w > 2 sec 2. Select Display Hold (“HLdon”) v repeatedly 3. Confirm setting Symbol “ ” on the display: application is active 4. If necessary: zero the balance/scale Display Enter CF w 5. Place sample on balance/scale 6.
Totalizing Purpose With this application program you can add up successive weight values exceeding the capacity of the balance/scale. Example: Totalizing weight values Step Key (or instruction) 1. Select application program w > 2 sec 2. Select Totalizing v repeatedly 3. Confirm setting Symbol “ ” on the display: application is active 4. If necessary: zero the balance/scale Enter CF w 5. Place sample on balance/scale (in this example, 380 g) 6. Store value in memory.
Specific Gravity Purpose Use this application program to determine the specific gravity of a sample. The result is displayed with one decimal place. Beaker and wire not included with balance/scale. Example: Determine the specific gravity of a solid. Step Key (or instruction) 1. Select application program w > 2 sec 2. Select Specific Gravity v repeatedly 3. Confirm setting Symbol “ ” on the display: application is active Enter CF 4. If necessary, zero the balance/scale w 5.
Calibration/Span Adjustment Calibration is recommended after initial installation and each time the balance/scale is moved. Features Calibration/adjustment can be performed only when: – there is no load on the balance/scale, – the balance/scale is zeroed, and – the internal signal is stable. If these conditions are not met, an error message is displayed. The weight required for calibration/ span adjustment is displayed.
Configuration (Setup Menu) To configure the balance/scale; i.e., adapt the balance/scale to individual requirements. Step Key (or instruction) 1. Switch off the balance/scale e 2.
Error Codes Error codes are shown on the main display for approx. 2 seconds. The program then returns automatically to the previous mode. Display/Problem No segments appear on the display oL UL disErr calErr appErr PrtErr Bal.Err Sys.Err Max.
Overview Specifications Model Weighing capacity Readability Tare range (subtractive) Linearity Operating temperature range Stabilization time (average) Adaptation to ambient conditions Calibration weight Net weight, approx. Pan size Power source/voltage/frequency Power consumption (average) Power requirements, DC voltage Hours of operation with battery: – Mignon, AA alkaline manganese (2200 mAh), avg. – Rechargeable 1500 mAh, fully ch., (NiMH), avg.
Accessories (Options) Product Data interface, mounting kit – RS-232 interface with cable – USB interface with cable – – – – Order No.
C 19
Sartorius AG Weender Landstrasse 94–108 37075 Goettingen, Germany Phone +49.551.308.0 Fax +49.551.308.3289 www.sartorius.com Copyright by Sartorius AG, Goettingen, Germany. All rights reserved. No part of this publication may be reprinted or translated in any form or by any means without the prior written permission of Sartorius AG. The status of the information, specifications and illustrations in this manual is indicated by the date given below.