Operating instructions
Practical Example
Determining an Unknown Piece Count; Weighing in the Preset Reference Sample Quantity
Settings (changes in the factory settings required for this example):
Setup: Application parameters: Application 1: Counting: Average piece weight updating: Manual
Setup: Printout: Application defined: Auto print on init.: All values
Step Key (or instruction) Display/Output
1. Delete previous setting if necessary c
2. Prepare a container for the Place the empty container
parts to be counted on the scale
3. Tare the scale t
4. Place the reference sample quantity Place the displayed number
on the scale (example: 10 pcs) of parts in the container
5. Determine the average piece weight
Start soft key
(number of decimal places displayed
depends on the scale model)
6. If necessary, increase the number Place additional parts
of parts and perform reference in the container
sample updating (example:
Update soft key
7 additional pieces)
7. Weigh uncounted parts Place parts to be counted
in the container
8. If desired, print total piece count p
(here: 153 pcs)
57
nRef + 10 pcs
wRef + 2.14800 g
nRef + 17 pcs
wRef + 2.15800 g
Qnt + 153 pcs