Operating instructions

Time-Controlled Functions K
Purpose
With this application program, you
can configure the scale to perform
certain functions (e.g., automatic print-
out of values, store value in totalization
memory) at a given time or after a set
interval.
You can use the “Time-Controlled
Functions“ application in combination
with a program chosen from Applica-
tion 1 (e.g., counting, weighing in
percent) and one from Application 3
(e.g., totalizing, formulation).
Available Features
Time-controlled activation of scale
functions:
one time only, at a given time
(
Setting= is displayed in the
text line)
repeatedly, at given intervals
(
Interval= is displayed in
the text line before the function is
started, and
Repeat= is displayed after the
function is started)
Functions that can be time-controlled
include:
Acoustic signal
Lock in readout
Automatic printout of values
Store values for totalizing,
formulation or statistics
Print time in addition to weight value
Store value independent of stability
Tare the scale after printout of weight
values
Press the corresponding soft key to
cancel time-controlled functions
Factory Settings
Function after time interval:
Automatic printout of
values
Automatic function restart: On
Storage mode:
Without stability
Print then tare: On
Soft Key Functions
Stop Stop the application
Quit Confirm performed function
(e.g., “Lock in readout” or
“Beep”)
Interv Store input interval for time-
controlled functions
Set. Store input time for one-time
performance of function
Printout for Time-Controlled Functions
If the “Automatic printout of values”
parameter is set, the time and weight
(or other value) are printed out.
Time: 10:15:00
N + 150.00 g
Time: Time that the values were
stored
N: Net weight
77
Preparation
Turn on the scale: press e
> The Sartorius logo is displayed
Select the “Time-controlled functions” application in the Setup menu: press H
Select the
Application parameters: press the q key 2+,
then the
O soft key
Select
Application 2 (control functions): press the q soft key,
then the
O soft key
Select
Time-controlled functions: press the Q or q soft key
Confirm
Time-controlled functions: press the O soft key
Time-controlled Function after Beep
functions time interval Lock in readout
o Automatic printout of values
Store value in applicat. 3 memory
Automatic function o On
restart Off
Storage mode o Without stability
After stability
After higher stability
Print then tare On
o Off
o = factory setting
see also the “Application Parameters (Overview)” in the chapter entitled “Configuration”
Save settings and exit the Setup menu: press the
oo soft key
Print Net Values without Printout of Time
Select the Setup menu: