Operating instructions
Time-Controlled Functions K
Purpose
With this application program, you
can configure the scale to perform
certain functions (e.g., automatic print-
out of values, store value in totalization
memory) at a given time or after a set
interval.
You can use the “Time-Controlled
Functions“ application in combination
with a program chosen from Applica-
tion 1 (e.g., counting, weighing in
percent) and one from Application 3
(e.g., totalizing, formulation).
Available Features
– Time-controlled activation of scale
functions:
– one time only, at a given time
(
Setting= is displayed in the
text line)
– repeatedly, at given intervals
(
Interval= is displayed in
the text line before the function is
started, and
Repeat= is displayed after the
function is started)
Functions that can be time-controlled
include:
– Acoustic signal
– Lock in readout
– Automatic printout of values
– Store values for totalizing,
formulation or statistics
– Print time in addition to weight value
– Store value independent of stability
– Tare the scale after printout of weight
values
– Press the corresponding soft key to
cancel time-controlled functions
Factory Settings
Function after time interval:
Automatic printout of
values
Automatic function restart: On
Storage mode:
Without stability
Print then tare: On
Soft Key Functions
Stop Stop the application
Quit Confirm performed function
(e.g., “Lock in readout” or
“Beep”)
Interv Store input interval for time-
controlled functions
Set. Store input time for one-time
performance of function
Printout for Time-Controlled Functions
If the “Automatic printout of values”
parameter is set, the time and weight
(or other value) are printed out.
Time: 10:15:00
N + 150.00 g
Time: Time that the values were
stored
N: Net weight
77
Preparation
● Turn on the scale: press e
> The Sartorius logo is displayed
● Select the “Time-controlled functions” application in the Setup menu: press H
● Select the
Application parameters: press the q key 2+,
then the
O soft key
● Select
Application 2 (control functions): press the q soft key,
then the
O soft key
● Select
Time-controlled functions: press the Q or q soft key
● Confirm
Time-controlled functions: press the O soft key
Time-controlled Function after Beep
functions time interval Lock in readout
o Automatic printout of values
Store value in applicat. 3 memory
Automatic function o On
restart Off
Storage mode o Without stability
After stability
After higher stability
Print then tare On
o Off
o = factory setting
see also the “Application Parameters (Overview)” in the chapter entitled “Configuration”
● Save settings and exit the Setup menu: press the
oo soft key
Print Net Values without Printout of Time
Select the Setup menu:
Setup: Printout: Application-defined output: Auto print upon initialization: Off