User manual

102
Specifications
User Manual for Secura, Quintix, Practum
Model-specific Data
Verified Models with Country-specific Type Approval Certificate
Secura
®
Model 224-1x
1
) 124-1x
1
) 613-1x
1
) 513-1x
1
) 313-1x
1
) 213-1x
1
) 6102-1x
1
) 5102-1x
1
) 3102-1x
1
)
Accuracy class
kkKKKKK K K
Type
2
) SQP-A SQP-A SQP-B SQP-B SQP-B SQP-B SQP-C SQP-C SQP-C
Max g 220 120 610 510 310 210 6,100 5,100 3,100
Scale interval d mg 0.1 0.1 1111101010
Verification scale interval e mg 1 1 10 10 10 10 100 100 100
Min mg 10 10 20 20 20 20 500 500 500
Tare (subtractive)
< 100% of max. weighing capacity
Minimum sample weight
as per USP, typical
g 0.12 0.12 1.5 1.5 1.5 1.5 12 12 12
Typical stabilization time s 2211111 1 1
isoCAL:
– Temperature change
– Time interval
K
h
1.5
4
1.5
4
2
6
2
6
2
6
2
6
2
6
2
6
2
6
Display result
(depending on the set filter level)
s 0.2 0.2 0.1/0.2 0.1/0.2 0.1/0.2 0.1/0.2 0.1/0.2 0.1/0.2 0.1/0.2
Weighing pan size mm d 90 d 90 d 120 d 120 d 120 d 120 d 180 d 180 d 180
Weighing chamber height mm 209 209 209 209 209 209
Net weight, approx. kg 5.1 5.1 5.1 5.1 5.1 5.1 5.2 5.2 5.2
Model 2102-1x
1
) 1102-1x
1
) 612-1x
1
)
Accuracy class
KKK
Type
2
) SQP-C SQP-C SQP-C
Max g 2,100 1,100 610
Scale interval d mg 10 10 10
Verification scale interval e mg 100 100 100
Min mg 500 500 500
Tare (subtractive)
< 100% of max. weighing capacity
Minimum sample weight
as per USP, typical
g121212
Typical stabilization time s 1 1 1
isoCAL:
– Temperature change
– Time interval
K
h
2
6
2
6
2
6
Display result
(depending on the set filter level)
s 0.1/0.2 0.1/0.2 0.1/0.2
Weighing pan size mm d 180 d 180 d 180
Weighing chamber height mm
Net weight, approx. kg 5.2 5.2 5.2