User manual
63
Checkweighing
User Manual for Secura, Quintix, Practum
Checkweighing
Purpose: Use this application to check whether a weight value falls within the specified
tolerances. This application also makes it easy to fill sample materials to a specified
target weight.
t Select the Menu key in any application.
t Select the Checkweighing symbol in the menu.
Max 2100 g d
=
0.01 g
0.00
Min g
0.00
Max g
0.00
y The Checkweighing application appears.
The set thresholds are displayed under Min and Max.
Min g
0.00
Max g
0.00
t To set the thresholds, select the gray button.
Checkweighing
Min Max
0.00
00.0
00.0
0.00
0.00
0.00
y The Checkweighing settings window appears.
t Select or enter the lower threshold (minimum) and the upper threshold
(maximum).
Select t, enter the desired value, and confirm with l.
t Select
l to confirm.
Max 2100 g d
=
0.01 g
0.00
Min g
190.00
Max g
195.00
y The set thresholds are displayed.
t Select A to zero the balance if necessary.
t Place a container on the weighing pan.