Calibration and Adjustment
User Manual for Secura, Quintix, Practum
Calibration/Adjustment using an Internal Calibration
Weight (Secura
®
and Quintix
®
only)
The balance is fitted with an internal calibration weight in the housing. This
calibration weight is automatically placed on the balance by a motor during
internal calibration and adjustment.
t Ensure that the balance is stable and the weighing pan is empty.
t Select A to zero the balance.
t Select the Menu key at the bottom left of the display in the current application.
ComponentsStatistics
MixingWeighing
y The menu appears.
t Select the CAL button to access the calibration function.
Calibrate balance
CAL-Extern
CAL-Intern
y The Calibrate balance window appears.
t Select CAL-Intern.
y Calibration begins.
The internal weight is automatically placed on the balance in the housing.
The balance is calibrated and then immediately adjusted.
The internal weight is automatically removed from the balance.
Date and Time: 14-Jun-2012 13:10
Calibration function: Internal calibration
Start calibration: Start: manually
Calibration deviation: Dev 0.01 g
Adjustment function: Internal adjustment
Adjustment deviation: Dev 0.00 g
Report
y A report appears on the display.
The report indicates the deviations identified during calibration.
The report also gives the result of the adjustment process.
t Close the report window: touch <.
y The balance is now calibrated and adjusted.