Calibration and Adjustment
User Manual for Secura, Quintix, Practum
Calibration/Adjustment using an Internal Calibration
Weight (Secura
®
and Quintix
®
only)
The balance is fitted with an internal calibration weight in the housing. This
calibration weight is automatically placed on the balance by a motor during
internal calibration and adjustment.
t Ensure that the balance is stable and the weighing pan is empty.
t Select A to zero the balance.
t Select the Menu key at the bottom left of the display in the current application.
ComponentsStatistics
MixingWeighing
y The menu appears.
t Select the CAL button to access the calibration function.
Calibrate balance
CAL-Extern
CAL-Intern
y The Calibrate balance window appears.
t Select CAL-Intern.
y Calibration begins.
– The internal weight is automatically placed on the balance in the housing.
– The balance is calibrated and then immediately adjusted.
– The internal weight is automatically removed from the balance.
Date and Time: 14-Jun-2012 13:10
Calibration function: Internal calibration
Start calibration: Start: manually
Calibration deviation: Dev 0.01 g
Adjustment function: Internal adjustment
Adjustment deviation: Dev 0.00 g
Report
y A report appears on the display.
The report indicates the deviations identified during calibration.
The report also gives the result of the adjustment process.
t Close the report window: touch <.
y The balance is now calibrated and adjusted.