USB Port
User Manual for Secura, Quintix, Practum
Interface Specification
Data Output
The data output can be activated either manually when a print command is
received, automatically synchronized with the display, or at defined intervals
(seeapplication programs and autoprint settings).
Data Output by Print Command The print command can be transmitted by pressing a or by a software command
(Esc P).
Automatic Data Output In autoprint mode, data is output to the data interface port without an extra print
command. Data can be output automatically synchronized to the display or at set
intervals, with or without stability.
If the automatic data output is activated in the Device Configuration, it starts
immediately after the balance is turned on.
Data Output Formats
You can output the values with or without an ID code. The output format is
configured in the device settings (see “Printout” on page31).
Example: Output without ID + 253 pcs (16 characters output with “SBI” only)
Example: Output with ID Qnt + 253 pcs (22 characters output – with printer and “PC text format”
always with ID)
Data Output Format with 16 Characters
Characters that are blank on the display are printed as spaces.
Display values without a decimal point are printed without a decimal point.
Normal Operation
Position 12345678910111213141516
+ AAAAAAA* EEECRLF
or -......... ***
or **********
* Space CR Carriage return
A Displayed characters LF Line feed
E Measurement unit symbol . Decimal point or comma