Operating Instructions Sartorius Signum® 1| Signum® 2 | Signum® 3 Models SIWABBP | SIWSBBP | SIWSBBS | SIWRDCP | SIWADCP | SIWSDCP | SIWSDCS | SIWAEDG Complete Scales 98648-014-73
Intended Use Intended Use Range of Models Signum® 1, 2 and 3 are precise and rugged complete scales that give you reliable weighing results. Three different type of weighing technology are used in the Signum® series, offering different performance levels: The Signum® series of compact scales includes models with strain-gauge weighing systems as well as versions equipped with monolithic technology, using the principle of electromagnetic force compensation.
Content Content Intended Use �������������������������������������������� 2 Contents ��������������������������������������������������� 3 Warnings and Safety Precautions �������� 4 Getting Started..........................................
Warning and Safety Instructions Warning and Safety Instructions Signum scales comply with the European Council Directives as well as international regulations and standards for electrical equipment, electromagnetic compatibility, and the stipulated safety requirements. § Read these operating instructions thoroughly before using your device. That way you will prevent damage to the equipment. ! Do not use this equipment in hazardous areas.
Getting Started Getting Started Unpacking the Equipment § After unpacking the equipment, please check it immediately for any external damage. $ If you detect any damage, proceed as directed in the chapter entitled “Care and Maintenance" under “Safety inspection." $ Save the box and all parts of the packaging for any future transport. Unplug all connected cables before packing the equipment.
Getting Started Getting Started Installing the Display and Control Unit The following options are available for installing the control unit: (the pictures depict the SIW*DCP model) § Fastening the display and control unit to the weighing platform: § Guide the display and control unit onto the retainer bracket. § Level the weighing platform (see page 7).
Getting Started § A recessed space is provided in the scale base, accessed from the bottom of the scale, for any excess length of cable. § Guide the connecting cable along the channel on the bottom of the weighing platform. § Use the cable clamps provided to affix the cable that connects the display and control unit to the weighing platform to the bottom of the column. § Turn the weighing platform right side up and place it so that it rests on its feet.
Getting Started Connecting the Device to AC Power $ Check the voltage rating and plug design $ The equipment is powered through the installed power cord. The power supply is integrated into the indicator. The device can be operated with a voltage of 100 V to 240 V. Make sure that the voltage rating printed on the manufacturer’s ID label is identical to that of your local line voltage.
General View of the Equipment Gerätedarstellung General View of the Equipment Signum® 1 Signum1 1 1 1 2 2 3 3 2 On Standby 0 4 Signum® 2 T Fn Tare Zero Function ISOTest 44 Tarataste 55 Tare key 6 Funktionstaste (Brutto/Netto 6 Function key (toggle between gross/ umschalten) net values) Justieren oder Kalibrieren starten 77 Start calibration or adjustment Print A 5 8 6 7 17 18 19 16 15 88 99 Signum2 10 10 1 2 CF On Standby n 3 Scale# REF Clear Function 0 OK Reference T
Operating Concept Bedienkonzept Tastatur Operating Concept Die Bedienung von Signum 1, Signum 3 ® Signum® 2 und Signum® 3 erfolgt mit Keypad wenigen Tasten. Dafür® 1, haben die ® 2 Operating the Signum Signum Tasten im Messbetrieb Menü ® 3 involvesund and Signum just im a few keys. unterschiedliche Funktionen. Einige These keys have one function during Tasten haben and darüber hinaus neben measurement another during ihrer normalen eine zweite Funktion configuration.
Operating Concept Signum® 2 and 3 only I Displays applications and manual tare values: Immediately closes information display: I press and hold longer than 2 seconds. The information is displayed in succession. Signum® 2 and 3 only c – Quits an application or deletes and input character Signum® 3 only 0, 1, 2... 9, . Enters numbers, letters and other characters Numeric Input via the Keypad (Signum® 3 only) § To enter numbers (one digit at a time): Press 0, 1, 2...
Operating Concept Bedienkonzept Operating Design Display Modes Display /Modes Anzeige Display Printing Drucken is active aktiv Scale weighWägebereiche ing dercapacity Waage Tare assignment Tarabelegung Bar graph Bargraph Measured value line Messwertzeile Application symbol Applikative Piktogramme Plus/minus sign; Vorzeichen/ stability symbol Stillstand Unit Speicher Battery Batteriesymbol betrieb Einheit Memory Calculated Verrechnete value Werte GMP printing GMP-Protokoll is active aktiv Re
h Operating Concept 1 2 3 4 5 6 7 20 19 8 9 18 10 Drucken Appl. 1 aktiv Appl. 2 Appl. 3 17 Verrechnete Wägebereiche Werte der Waage Display in Weighing Mode Tarabelegung The illustration above depicts all ofBatteriethe 16 15 14 13 12 11. Numeric display; e.g., showing reference value (Signum® 2 and 3 only) 11 main display elements and symbolsbetrieb that can be shownEinheit during weighing. (Signum® 2 and 3 only): GMP-Protokoll 12. Symbol indicating data transfer aktiv 1.
Operating Concept 1 4 2 Setup Menu Display: Text Menu (Example) Menu Operating Design 5 6 7 3 4 Setup Menu Display: Numeric operating menu (example) Alphanumeric Input in the Menu The keys below the display are used to navigate and make entries in the device menu. Numeric input in the Signum® 3 operating menu Enter values (date and time, etc.
Operating Concept Error messages – If a key is inactive, “-------” and/ or “No function” is displayed briefly (2 sec.) – Temporary errors are displayed for 2 seconds in the measured value/result line via an error code (e.g., Inf 09); fatal errors (e.g., Err 101) are displayed continuously until “Reset" is used. Error codes are described in detail under “Error Codes” on page 92. Data Output Backups Printer You can connect two strip or label printers to Signum® 1, 2 or 3.
Settings Settings – – – – – You can configure the Signum by selecting parameters in the Setup menu. These are divided into the following groups (menu level 1): Application parameters Fn key function Device parameters Device-specific information “Info" Language Setting the Language Example: Selecting “US Mode" for the language e ) While all segments are lit, press the ) key When used in legal metrology, not all parameters can be accessed.
Settings Protecting Setup With a User Password Example: Assign a password (in this example, AB2) to protect the application program settings APPL and the device parameters SETUP from unauthorized changes e ) 1.) Turn on the Signum® 2.) W hile all segments are lit, press ) p, p, p ) 9.) E nter the second character using the p and k keys (in this example: B) 10.) Save the character The first item in the Main menu is shown: APPL k ) k 3.
Datentransfer Settings verrechneter en folgende Symbole: raph mit 10%-Marken mum »Kontrollwägen« Taraspeicherbelegung, verrechnete Werte, Kennzeichnung der aktiven Wägeplattform pt Hinweis auf Tara-Handeingabe (über Barcodeleser) bei TaraInformationen Operating Menu Overview wert »Kontrollwägen« wp1 Anzeige der aktiven Wägeplattform Anschluss von 2 WägeplattIn the Setup menu, you canbei configure Menu levels are identified by texts, and numeric codes identify the individual settings.
Settings Operating Menu = Setting/Function available for Signum® 1 and 2 only = Setting/Function available for Signum® 2 and 3 only * Factory setting Menu appl appl1 1) Application programs W / weigh. 3.5. 3. 7. 3. 25. 9. 1. Z 3.5.1 * 3.5.2 3.5.3 3.5.4 3.5.5 3.5.6 3.5.7 3.5.8 3.5.9 3.5.10 3.7.1 * 3.7.2 3.25.1 3.25.2 9.1.1 * 9.1.2 * / count. 3.5. 3.6. 3.7. 3.8. 3.9. 3.11. 3.12. 3.13. 3.25. 9.1.
Settings appl appl 1 1) Znm / neutr.m Minimum load for automatic taring and automatic printing2) Numeric menu similar to Counting 3.6. Minimum load for initialization Numeric menu similar to Counting 3.7. Automatic taring: first weight tared2) Numeric menu similar to Counting 3.8. Automatic start of applications when you switch on the device with most recently saved application data2) Numeric menu similar to Counting 3.9. 3.10. 3.11. 3.13. 3.25. 9.1. 3.9.1 * 3.9.2 3.9.3 3.9.
Settings appl appl 1 1) L / perc.wg Minimum load for automatic taring and automatic printing 2) Numeric menu similar to Counting 3.6. Minimum load for initialization Numeric menu similar to Counting 3.7. Automatic taring: first weight tared 2) Numeric menu similar to Counting 3.8. Automatic start of applications when you switch on the device with most recently saved application data 2) Numeric menu similar to Counting 3.9. 3.10. 3.11. 3.13. 3.15. 3.25. 9.1. 3.9.1 * 3.9.2 3.9.3 3.9.
form chen r D für Wägewert oder verrechWert, Settings ür Nullstellung oder bei tarierter lattform (nur bei geeichten en) ertzeile lung der Wägewerte oder des verten Werts durch Ziffern und aben. appl t und Stillstand llstand der Waage werden die 2 1) appl htseinheit oder die Einheit für verrechneten Wert angezeigt.
Settings totaliz Totalizing 3.5. Minimum load for automatic taring and automatic printing2) Numeric menu similar to Counting 3.6. Minimum load for initialization Numeric menu similar to Counting 3.7. Automatic taring: first weight tared2) Numeric menu similar to Counting 3.8. Automatic start of applications when you switch on the device with most recently saved application data2) Numeric menu similar to Counting 3.16. 3.17. 3.22. 3.23. 3.25. 9.1. 3.16.1 * 3.16.2 3.17.1 3.17.2 * 3.22.
Settings appl a.tare 1) 3. 7. Automatic taring: first weight tared Off On 3.5.1 * 3.5.2 3.5.3 3.5.4 3.5.5 3.5.6 3.5.7 3.5.8 3.5.9 3.5.10 Minimum load for automatic taring and automatic printing 1 digit 2 digits 5 digits 10 digits 20 digits 50 digits 100 digits 200 digits 500 digits 1000 digits 3.8.1 3.8.2 * utomatic start of applications when you switch on the device with most recently A saved application data Automatic (On) Manual (Off) 3.24.1 * 3.24.
Settings Key assignment k fn-key OFF * gronet 2. unit res 10 sqmin Signum® 1 only: Signum® 1 only: k key Gross/Net toggling Display of 2nd weight unit 10-fold increased resolution, display max. 5 seconds Display the minimum permissible sample quantity 3) Device Parameters Setup A password prompt is displayed if a password is configured 1 Weighing platform 1 wp-1 (Display designation of this menu level: 1 ) 1.1. 1.2. 1.3. 1.4. 1.5. 1.6. 1.7. 1.8. 1.1.1 1.1.2 * 1.1.3 1.1.
Settings Setup wp-1 1 1.9. 1.10. 1.11. 1.12. 1.13. 1.15. 1.16. 1.17. 1.18. 3. 1. 3.2. Calibration and Adjustment Ext. calibration/adjustment; default weight External calibration/adjustment; weight can be selected under menu item 1.18.1 Internal calibration/adjustment (models with built-in motorized calibration weight only) Setting the preload Deleting the preload 1.9.10 No function when you press the J key 1.10.1 1.10.
Settings Setup Interface port 1 Com1 2 (Display designation of this menu level: 2) off* WP2 Off Weighing platform 2 rs-232 * RS-232 sbi-std SBI standard version sbi-app SBI trade version (for legal metrology) Xbpi-232 * XBPI-232 1) 1.1. to 1.8. Numeric menu similar to WP1 1.9. 1.9.1 * 1.9.3 1.9.4 Calibration and Adjustment Ext. calibration/adjustment; default weight External calibration/adjustment; weight can be selected under menu item 1.18.1 Internal cal.
Settings Setup Com1 2 datProt sbi* 6.3. 7.2. 9.1. xbpi-232 28 Signum 1-2-3 6.3.1 * 6.3.2 6.3.4 6.3.7 Time-dependent automatic data output 1 display update 2 display updates 10 display updates 100 display updates 7.2.1 7.2.2 * Data output: Line format For raw data: 16 characters For other applications: 22 characters 9.1.1 9.1.
Settings Setup Com1 2 datProt SMA interface function SMA 5.1. 5. 1. 1 5. 1. 2 5. 1. 3 5. 1. 4 5. 1. 5 5. 1. 6 5. 1. 7 * 5. 1. 8 Baud rate 150 baud 300 baud 600 baud 1200 baud 2400 baud 4800 baud 9600 baud 19200 baud 5.2. to 5.6. Numeric menu similar to SBI Printer YDP01IS YDP01IS Strip printer Label printer Label printer with manual feed line* label lab ff YDP02 variants YDP02 5.1. 5.2. 5.3. 5.4. Printer configuration 5.1.4 * 5.1.5 5.1.6 5.1.
Settings Setup Com1 2 Printer Universal interface Uni-pri 5.1. 5.2. 5.3. 5.4. 5.6. Baud rate 150 baud 300 baud 600 baud 1200 baud 2400 baud 4800 baud 9600 baud 19200 baud 5.2.2 5.2.3 5.2.4 5.2.5 * Parity Space 1) Odd Even None 2) 5.3.1 * 5.3.2 Number of stop bits 1 stop bit 2 stop bits 5.4.1 * 5.4.3 Handshake mode Software handshake Hardware handshake, 1 character after CTS 5.6.1 5.6.2 * Number of data bits 7 data bits 8 data bits YDP04IS * 5.1.1 5.1.2 5.1.
Settings Setup unicom 3 Interface 2 (optional) (Display designation of this menu level: 3 ) off* Off WP2 Weighing platform 2 RS-232 rs-232 sbi-std SBI standard version sbi-app SBI trade version (for legal metrology) XBpi-232 * XBPI-232 1) 1.1. to 1.8. Numeric menu similar to WP1 1.9. 1.9.1 * 1.9.3 1.9.4 1.9.10 Calibration and Adjustment Ext. calibration/adjustment; default weight Extern calibration/adjustment; weight can be selected under menu item 1.18.1 Internal cal.
Settings Setup unic om 3 Interface port 2 (optional) (Display designation of this menu level: 3 ) datProt Data protocol sbi * SBI standard version 5. 1. to 9.1. Numeric menu similar to COM1 XBPI-232 xbpi-232 xbpi-485 sma Profibus XBPI-485 Network address: Selectable from 0 to 31 0 to 31 SMA interface function 5. 1. to 5.6. Numeric menu similar to COM1 0 to 126 Profibus Address: Selectable from 0 to 126 SRC. IP src.name lis.port subnet gate-ip DEST. IP dest.
Settings Setup uniCom1 3 printer YDP01IS YDP01IS Printer configuration line * label lab ff Strip printer Label printer Label printer with manual feed YDP02 variants YDP02 5.1. to 5.4. Numeric menu similar to COM1 YDP03-0CE, YDP20-0CE YDP03 5.1. to 5.4. Numeric menu similar to COM1 YDP023IS YDP02IS line * label Strip printer Label printer Universal interface Uni-pri 5.1. to 5.6.
Settings Setup External key ctrl io 5 (Display designation of this menu level: 5 ) Input ports ctr inp 8. 4. 8.17. 8.18. 8.19. 8.20. 8.21. 8.4.4 8.4.5 8.4.6 8.4.7 8.4.8 8.4.9 8.4.10 8.4.11 Trigger J key function Trigger k key function Trigger n key function Trigger O key function Combined zero/tare function Trigger ( key function Trigger e key function Trigger c key function 8.4.12 8 4.13 Trigger I key function Trigger D key function 8.17.1 ’’ 8.17.
Settings Setup barcode 6 Barcode (Display designation of this menu level: 6 ) ref * tare ID1 1) Input Header ext.keyb Store value as reference Store value as tare Save as ID 1 Enter value on display (triggered when a key is pressed) Store value as tare or ID code, depending on barcode header External computer keypad prtprot 7 Printouts (Display designation of this menu level: 7 ) 7. 4. 1 (blank) 7. 4. 2 (blank) 7. 4. 3 (ID 1) 7. 4. 4 (ID 2) 7. 4. 5 (ID 3) 7. 4.
Settings Setup prtprot 7 7.10. Configuration list, individual, interface 2 blank ———------form-fe dat/tim 2) time 2) glphead glpfoot trans.no id1 1) id2 1) id3 1) id4 1) net gross tare tare2 headr. 1 headr. 2 sn scale 7.11. Configuration list, components, interface 2 blank ’’ sn scale 7.12. Configuration list, total, interface 2 blank ------form-fe dat/tim 2) time 2) glphead glpfoot trans.no id1 1) id2 1) id3 1) id4 1) headr. 1 headr. 2 sn scale 7.13. 7.14. 7.15. 7.16. 7.17. 7.18. 9.1.
Settings Setup utilit 8 Operation (Display designation of this menu level: 8 ) 8. 3. 8. 7. 8. 8. 8.9. 8.3.1 * 8.3.2 8.3.3 8.3.4 8.3.5 8.3.6 8.3.7 8.3.8 8.3.9 8.3.10 8.3.11 8.3.12 8.3.13 8.3.14 8.3.15 8.3.16 8.3.17 8.3.18 8.3.19 Keypad All available All blocked Numeric keypad Toggle weighing platform Zero Tare FN isoTEST Print x10 Toggle gross/net CF Ref OK Toggle Info ’D’ ID Mem 8.7.1 8.7.2 * Automatic shutoff of display and control unit Automatic shutoff acc. to menu item 8 9.
Settings Setup time 3) Time Format: e.g. 10.07.41 (hours.minutes.seconds) date 3) Date Format: 01.05.02 (day.month.year); U.S. mode: (month.day.year) User password u-code S et, change and delete password here (max. 8 characters): e.g. 12345678 sqmin 9 (sqmin-s: For service only: Enter the min. sample quantity) Display No * Yes Print in GLP header No * Yes Device-specific information Info Service information Service 10.04.
Operation Operation Weighing Weighing W The basic weighing function is always accessible and can be used alone or in combination with application programs, such as Counting, Checkweighing, Weighing in Percent, etc.
Operation Example with Signum® 1: Switch on the device, zero the scale, tare the container weight, place sample in the container, toggle display to gross weight or to second weight unit or 10-fold resolution, print results. e isplay with tared scale and D filled container 1.) Switch on the scale ll display elements appear A for approx. 1 second (display check) Signum® 1: k Display with no load on scale ( Signum® 2 and 3: L t he gross weight (in this example, 50 g for container + 120.
Operation Example with Signum® 1: Tare the scale by placing a container on the weighing platform Example with Signum® 3: Numeric input of tare weight, print the results e e 1.) Switch on the scale. The automatic self-test runs. When the weight readout is shown, the scale is ready to operate and automatically set to zero. With no load on the platform, you can zero the weighing platform at any time by pressing (. 250 2.
Operation Example with Signum® 3: Weighing with variable tare values, printing results, deleting tare values e 1.) Switch on the scale. The automatic self-test runs. When the weight readout is shown, the scale is ready to operate and automatically set to zero. With no load on the platform, you can zero the weighing platform at any time by pressing (. 2.) Place empty container on the platform. 3.) Tare the scale Note: If the automatic tare function is enabled, you do not need to press the ) key.
Operation Calibration and Adjustment Switch Purpose Perform calibration to determine the difference between the value displayed and the actual weight on the platform. Calibration does not entail making any changes within the weighing equipment. During adjustment, the difference between the measured value displayed and the true weight of a sample is corrected, or is reduced to an allowable level within maximum permissible error limits.
Operation SIWR, SIWA models: Geographical data – Elevation and latitude or gravitational acceleration displayed after Cal is shown when the Signum is switched on, if these values are supported by the weighing encoder: setup utilit 8.12.: (show geographical data before calibration/adjustment) For each of these parameters, the term is displayed first (Altitud, Latitud or Gravity) for 1 second, and then the corresponding value is displayed continuously until you press ).
Operation § Open the Setup menu: Press the ) key Preparation § Switch on the scale: Press the e key § While all segments are lit, press the ) key § Select the Setup menu: Press the k key repeatedly until Setup is displayed $ Select weighing platform 1 “WP1:" Press the ) key or $ Select interface 1 “COM-1" or interface 2 “unicom" (depending on the interface): Press the k key Select weighing platform 2 “WP2:" Press the ) key Setup wp-1 1 1.9. 1.10. 1.11. 1.12. 1.13. 1.15. 1.16. 1.17. 1.18. 1.9.
Operation Example: External calibration and manual adjustment with default weights (weighing parameters: factory settings) ( 1.) Zero the scale. J he difference between the T measured value and the true weight of the sample will be displayed with plus/minus signs. 2.) Start calibration (e.g., when adjustment prompt flashes WP). c.ext.def appears for two seconds. You are prompted to place the required weight on the platform (e.g., 10,000 g). Ext.
Operation SQmin Function Purpose To display the allowable minimum sample quantity “SQmin” (sample quantity minimum) in accordance with the United States Pharmacopoeia (USP). According to USP guidelines, the uncertainty of measurement may not exceed 0.1 % of the sample quantity when substances are weighed with the highest degree of accuracy for volume determination.
Operation Example Determining sample weights while monitoring the minimum sample quantity (in this example, SQmin = 100 g) Settings (different from the factory settings): Setup: Equipment: SQmin: Display: on Step Press key (or action) 1. Switch on the scale and enter settings as above e 2. Place the container for the sample on the scale and tare ) 3. Measure the weight of a sample (in this example: minimum sample quantity not reached) Place the sample on the scale 4. Print weight value p 5.
Operation Individual ID codes (identifiers) Signum® 3 only: You can assign codes (such as product name, batch number, etc.) for identification of measured values on printouts. Factory settings for the ID code names ID1: ID2: ID3: ID4: ID1 ID2 ID3 ID4 Factory settings for the ID code values No default values set. Characteristics – Assign up to four ID codes. – Assign both a name and a value to each ID code.
Operation Example with Signum® 3: Enter ID code names. “Batch number" and “Customer" should be entered for ID 1 and ID 2. e ) 10) Save the character 1) Switch on the scale 11) E nter additional letters as described above.
Anwendungen kombinieren Combining Application Programs Anwendungsprogramme Applikationen 1 - 3: Übersicht Signum1 Application Programs Applications 1 - 3: Overview Tastatur Keys Anzeige Display Signum® 1 Signum® 2 Signum® 3 6 Tasten Signum® 1 14 Tasten Signum® 2 17 Tasten plus Signum® 3 numerische Tastatur 17 keys plus numeric keypad 14-Segment plus 14-segment Applikations plus Symbole application symbols 6 keys 14 keys 14-Segment 14-Segment 14-segment plus 14-segment Applikations plus Symbol
Counting Application Z Counting Application Z With the Counting application, you can determine the number of parts which each have approximately equal weight.
Counting Application Z Preparation § Switch on the scale: Press the e key § While all segments are lit, press the ) key § Select the Setup menu: Press the k key repeatedly until appl is displayed § Confirm appl: Press the ) key § Select the Counting application: Press the k key repeatedly and confirm with the ) key Counting application parameters 3.6. Minimum load for initialization 3.6.1* 1 digit 3.6.2 2 digits 3.6.3 5 digits 3.6.4 10 digits 3.6.5 20 digits 3.6.6 50 digits 3.6.7 100 digits 3.6.
Counting Application Z Average piece weight updating Reference In Setup, Sample under: Updating InAppl: the Setup Z: menu: 3.12. ® 3: Appl: Z: 3.12. Signum ® 3 Signum appl1: count: 3.12. appl1: count : 3.12. You can define whether or not the You can define whether or is not the reference sample weight updated reference sampleduring weightweighing. is updated automatically automatically weighing.
Counting Application Z Example: Determining the number of uncounted parts. Settings (different from the factory settings): Setup: Application 1: Counting Setup: Device Parameters: Printout: PRTPROT: 7.6, then select the menu line items of your choice (see “Configuration” for options) r 5.) Start calculation of the reference sample weight 1.) Place empty container on the platform ) S ignum® 2: Set the number of reference parts using r: 1, 2, 5, 10, 20, etc.
Neutral Measurement Application Z nM Neutral Measurement Application Z nM With this application you, can use your weighing platform to measure the length, surface and volume of parts. The o symbol is displayed as the weight unit.
Neutral Measurement Application Z Parameter for saving weight values The reference weight is saved when the scale has stabilized. “Stability” is defined as the point at which the fluctuation of a measured value lies within a defined tolerance range. The narrower the tolerance range, the more stable the platform is at “stability”. In Setup, under: Appl: Z nM: 3.11. Signum® 3: appl1: Neutr.M:3.11.
Neutral Measurement Application Z nM Example: Measuring 25 m of electrical cable. Settings (different from the factory settings): Setup: Application 1: Neutral measurement Setup: Printout; PRTPROT 7.6. then select the menu line items of your choice 1.) Place empty container on the platform ) 2.
Averaging (Animal Weighing) Application V Averaging (Animal Weighing) Application V With the Averaging application, you can use your weighing platform for calculating weights as the average of a number of individual weighing operations. These individual operations are also known as “subweighing operations.” This function is used to determine weights under unstable ambient conditions or for weighing unstable samples (such as live animals). Characteristics – Averaging started manually or auto matically.
Averaging (Animal Weighing) Application V Preparation § Switch on the scale: Press the e key § While all segments are lit, press the ) key § Select the Setup menu: Press the k key repeatedly until appl is displayed § Confirm appl: Press the ) key § Select the Animal Weighing application: Press the k key repeatedly and confirm with the ) key Application Parameters: Animal Weighing 3. 6. Minimum load for start 3.6.1* 1 digit 3.6.2 2 digits 3.6.3 5 digits 3.6.4 10 digits 3.6.5 20 digits 3.6.6 50 digits 3.6.
Averaging (Animal Weighing) Application V Example: Measuring the weight of one mouse. Settings (different from the factory settings): Setup: Application 1: Animal weighing Setup: Printout; PRTPROT 7.6 then select the menu line items of your choice he averaging routine does T not begin until the fluctuation in weight value remains below a defined threshold over three consecutive measurements. The number of subweighing operations remaining is shown in the numeric display.
Weighing in Percent Application L Weighing in Percent Application L With the Weighing in Percent application, you can use your weighing platform to obtain weight readouts in percent which are in proportion to a reference weight. L is displayed as the weight unit.
Weighing in Percent Application L Parameter for saving weight values The reference weight is saved when the scale has stabilized. “Stability” is defined as the point at which the fluctuation of a measured value lies within a defined tolerance range. The narrower the tolerance range, the more stable the platform is at “stability”. In Setup, under: Appl: L: 3.11. Signum® 3: Appl1: Perc.wg: 3.11.
Weighing in Percent Application L Example: Weighing in 100% of a sample material. Settings (different from the factory settings): Setup: Application 1: Weighing in Percent Setup: Printout, PRTPROT 7.6, then select the menu line items of your choice If the weight is too light, an error code is shown in the main display: Inf 29 1.) Place empty container on the platform ) educe the minimum load R setting 5.
Checkweighing Application O Checkweighing Application O With the Checkweighing application, you can check whether the sample on the weighing platform matches a target value or lies within a given tolerance range. Checkweighing also makes it easy to fill sample materials to a specified target weight. Characteristics Signum® 3 only: – Enter the nominal or target weight (set point) and the tolerance range delimiters either using the keypad or by saving the weight value of a load on the platform.
Application 2: Checkweighing O Minimum Load The minimum load required for automatic taring of the container weight on the platform (first weight), or forfor automatic printout of of results, is or automatic printout results, configured in in Setup under: is configured Setup under: O: 3.5. 3.5. Appl: O:: Signum® 3: Appl: m.weigh: 3.5.
Application 2: Checkweighing O Example: 1 Checkweighing samples with a target weight of 1250 g and a tolerance range from -10 g to +30 g (using the factory-set application parameters) Settings (different from the factory settings): Setup: Application 2: Checkweighing Setup: Printout, PRTPROT: 7.6, then select the menu line items of your choice 1280 O O 1.) Begin input of target and tolerance values 2.) Place a sample with the target weight (in this example, 1250 g) on the platform 1240 O 3.
Checkweighing Toward Zero Application Checkweighing Toward Zero Application Example: 2 Checkweighing samples with a target weight of 1250 g and a tolerance range from -10 g to +30 g (using the factory-set application parameters) Settings (different from the factory settings): Setup: Application 2: Checkweighing toward zero (parameter 4. 7. 2) Setup: Printout, PRTPROT 7.6, then select the menu line items of your choice 1280 O O 1.) Begin input of target and tolerance values 2.
Classification Application Classification Application W With the Classification application, you can determine whether the weight of a given sample lies within the limits of a defined weight class. Characteristics – Classification with 3 or 5 weight classes. Configured in Setup under: Appl: W: 4.8. Signum® 3: Appl 2: class: 4.8.
Classification Application W Minimum Minimumload Load The Theminimum minimumload loadfor forthe thefirst firstclass classisis configured configuredininSetup Setupunder: under: Appl: Appl:W W: :3.6. 3.6. Appl Appl22: : ® Signum Signum®3:3: class class: :3.6. 3.6. Once Oncethe thelimit limitisisexceeded exceededby bythe theload, load, initialization initializationcan canbegin. begin.
Classification Application W Example: Defining three classes. Settings (different from the factory settings): Setup: Application 2: Classification Setup: Printout, PRTPROT 7.6, then select the menu line items of your choice O 1.) Begin input of class delimiters 6.) Place the sample on the weighing platform 110 O 2.) For Signum® 3 only: Enter the upper limit for Class 1 using the keypad (in this example, 110 g) 3.) Save the upper limit for Class 1 Read the result p + 130 O 4.
Totalizing Application s Totalizing Application s With the Totalizing application, you can add weight values to the totalizing memory. In addition to weight values, the number of separate values added to memory is also saved (transaction counter). Characteristics – Totalize up to 999 individual weights Save values automatically: – Simultaneous saving of net values and calculated values (if available). Configured in Setup under: Appl:s: 3.16. Signum® 3: Appl 3: totaliz: 3.16.
Totalizing Application s Preparation § Switch on the scale: Press the e key § While all segments are lit, press the ) key § Select the Setup menu: Press the k key repeatedly until appl is displayed § Confirm appl: Press the ) key § Select the Totalizing application: Press the k key repeatedly and confirm with the ) key Totalizing Application Parameters 3. 6. Minimum load for autosave 3.6.1* 1 digit 3.6.2 2 digits 3.6.3 5 digits 3.6.4 10 digits 3.6.5 20 digits 3.6.6 50 digits 3.6.7 100 digits 3.6.
Totalizing Application s Example: Totalizing weight values Settings (different from the factory settings): Setup: Application 3: Totalizing Setup: Printout, PRTPROT 7.6 Setup: Device Parameters: Printout: Printer 1: Select “Component log: Auto printout”, then select the menu line items of your choice PRTPROT: 7.7. Setup: Printout: Printer 1: Select “Total data record: Printout as per CF key,” then select the menu line items of your choice PRTPROT: 7.8. 4.
Net-total Formulation Application R Net-total Formulation Application R With this application, you can weigh in different components up to a defined total. Each component is saved in the net-total memory. Characteristics – Weigh in up to 999 components in series – Automatic taring of container weight. Configured in Setup under: Appl: R: 3.7. Signum® 3: Appl: a.tare: 3.7. – Restore factory default settings. Configured in Setup under: Appl: R: 9.1. Signum® 3: Appl: def.app: 9.1.
Net-total Formulation Application R Preparation § Switch on the scale: Press the e key § While all segments are lit, press the ) key § Select the Setup menu: Press the k key repeatedly until appl is displayed § Confirm appl: Press the ) key § Selecting the Net-total Formulation Application: Press the k key repeatedly and confirm with the ) key Net-total Formulation Application Parameters 3.6. Minimum load for autosave 3.6.1* 1 digit 3.6.2 2 digits 3.6.3 5 digits 3.6.4 10 digits 3.6.5 20 digits 3.6.
Net-total Formulation Application R Example: Weighing in 3 components of a formulation recipe. Settings (different from the factory settings): Setup: Application 3: Net-total Formulation Setup: Printout, PRTPROT: 7.7. Printer 1: Select “Component log: Auto printout”, then select the menu line items of your choice Setup: Printout, PRTPROT: 7.8 Printer 1: Select “Total data record: Printout as per CF key,” then select the menu line items of your choice 1.) Place empty container on the platform ) 2.
Net-total Formulation Application R he weighing platform is tared T and the component counter value is increased by one. Prompt to fill and save the second component is shown 8.) A dd the third component to the container until the desired target is reached (in this example, 2000 g). 5.) Add the second component to the container (in this example, 525 g) The weight of the second component is displayed The total weight is displayed O 9.) S ave the third component weight Cmp003+ 0.
Examples of Application Combinations in Signum® 3 Example: “Portioning” (counting, checkweighing with totalizing) Settings (different from the factory settings): Setup: Application parameters: Application 1: Counting (count) Setup: Application parameters: Application 2: Checkweighing (check.wg) Setup: Application parameters: Application 3: Totalizing: Saved value: Net + Calculated (3.23.3) Setup: Application parameters: Application 3: Totalizing: Autosave: On (3.16.
Examples of Application Combinations in Signum® 3 If the weight is too light, an error code is shown in the main display: Inf 29 educe the minimum load R setting or increase the reference sample quantity setting and the number of parts in the container D 11.) A dd desired number of pieces 7.) Toggle to Checkweighing saved automatically O 8.) Initialize Checkweighing 100 9.
Configuring Printouts Configuring Printouts Purpose You can configure individual printout formats for each application. Using the total data record for Totalizing and Net-total Formulation applications, you can define which parameters are printed using the c key. In the “Print parameters” Setup menu, single, component and total data records can be configured, which contain the available print items for the respective applications.
Configuring Printouts Example: Standard printout for data output from the Counting application Settings (different from the factory settings): Applications: Application 1: Counting Then access Setup: Printout: Printer 1: “Individual: print by pressing p” § § § § Select Setup: Press the k key repeatedly until setup is displayed Select Config. printout: Press the ) key Press the k key repeatedly until prtprot is displayed Press the ) key ) 1.) Press the ) key until 7.4 appears in the display k 2.
Product Data Memory for Signum® 3 Product Data Memory for Signum® 3 Purpose The product data memory stores initialization data and user data (product and tare values). Characteristics – The product data memory has sufficient capacity for 100 product or tare values. – For example, you can store 80 sets of application data and 20 tare values. – Each memory cell is uniquely identified by a number up to three digits.
Data Interface Data Interface For COM1 (25-pin D-Sub female connector), for model SIWSDCS/SIWSBBS with round socket* COM1 Standard: RS-232 SBI | XBPI protocol All Signum® models PC with serial RS-232 input port Printer: YDP04IS YDP02IS YDP12IS YDP20-0CE External Alibi memory: YAM01IS Option: RS-232 Option A31: “Clock” Option: RS-232 Option A32: “connectivity” Universal in: print/ tare Signum® 2 and 3 only Red-green-red display YRD14Z (uses digital control lines) Digital control lines (TTL/ 5V) <;=;set;>
Data Interface Pin Assignment Chart Pin Assignment Chart COM1 and UniCOM female connectors: Round socket with screw lock hardware for model SIWSDCS/SIWSBBS, stainless steel version. COM1 female connectors: 25-pin D-Submini female connector DB25S for model: SIW*DCP*..
Data Interface Cabling Diagram (Adapter Cable for PC) Indicator PC V24 Waage 25-pin Stecker Buchse Computer 9 pin TxD 2 2 RxD 3 3 CTS 5 4 DTR 20 8 GND 4/7 6 GND 14 5 Waage 25-pin Stecker Buchse Computer 25 pin TxD 2 3 RxD 3 2 CTS 5 20 DTR 20 5 GND 4/7 6 GND 14 7 Cable type: AWG 24 specification 86 Signum 1-2-3
Data Interface Configuring the Data Interface as a COM Port (datprot) You can configure the interface as a COM port in either COM1 or UniCOM, “Data Protocol” (datprot) menu item. SBI communication This is a simple ASCII interface. Data output is configured under menu items 6.1 and 6.3: – Manual output of displayed value with or without stability (menu items 6.1.1 and 6.1.2) – Automatic output of displayed value with or without stability (menu items 6.1.4 and 6.1.5) at intervals defined by display updates.
Data Interface Data Output Format Error Message: Each line in a print job can contain up to 22 characters (up to 20 printable characters plus two control characters). The first 6 characters, called the “data header”, identify the subsequent value. You can suppress the header under menu item 7.2 in the “Printouts” menu; in this case, the print job has up to 16 characters (up to 14 printable characters plus two control characters). Pos.
Data Interface Error Message: Example (output of weight value of +1255.7 g): Pos. 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 Pos. 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 S t a t * * * * * E r r * * # # * * * * CR LF G # * * * * + * * * 1 2 5 5 . 7 * g * * CR LF or *: #: S t a t * * * * * E r r * # # # * * * * CR LF Space Error code number (2 or 3 digits) ID Code Characters: Character K G# N T T2 Diff Targ. Nom. nRef pRef wRef Qnt mDef x-Net Setp Diff.W.
Data Interface Automatic Data Output (SBI) You can have the weight readout printed automatically1). This printout can be generated after a certain number of display updates2). You can also configure whether or not the auto-print function is dependent on the stability of the scale3). The display update frequency depends on both the scale model and the operating status. Examples: N + Stat Stat Stat 153.
Data Interface GMP-compliant Printouts When the corresponding menu item is active, the printout is bracketed by a GMP header and a GMP footer (GMP: “Good Manufacturing Practice”). The GMP header precedes the first measured result. The GMP footer is printed after the last result in a series of measurements (“GMP-compliant printout always for several application results,” menu item 7.13.3). To end a series of measured results, press and hold the p key.
Error Messages Error Messages Error messages are shown in the main display. Error messages are shown permanently. Information messages are shown for 2 seconds. After this the program automatically returns to the weighing mode.
Care and Maintenance / Disposal Care and Maintenance Service Regular servicing by a Sartorius technician will extend the service life of your equipment and ensure its continued weighing accuracy. Sartorius offers its customers service contracts with regular maintenance intervals ranging from 1 month to 2 years. Disposal Corrosive Environment $ Remove all traces of corrosive substances from the scale on a regular basis. Replacing the dust cover > Replace damaged dust cover.
Specifications General Specifications Digital protective interface According to EN45501 Data Interface Bidirectional RS-232C interface with control output (standard equipment) Additional data interface Optional Display 14-segment backlit display Weighing platform Housing: Display unit Dust and water protection acc.
Specifications Model-specific Data Weighing Data Signum® Regular SIWRDCP-1,-2,-3 Non-verifiable models Weighing capacity (kg) Readability (g) Resolution code 3 0.1 -I 3/6 1/2 -N 6 1 -R 6 0.5 -L 6 0.2 -I Calibration weight value (in grams) 2000 Accuracy class M1 5000 M2 5000 M2 5000 M1 5000 M1 Verifiable/verified models Type of construction EC type approval no.
Specifications Model-specific Data Weighing Data Signum® Regular SIWRDCP-1,-2,-3 Non-verifiable models Weighing capacity (kg) Readability (g) Resolution code Calibration weight (in grams) Accuracy class 15/35 5/10 -N 10000 M2 15/35 5/10 -M 10000 M2 35 5 -R 10000 M2 35 2 -L 10000 M1 35 1 -I 10000 M1 60 10/20 -M 60000 M1 Verifiable/verified models Type of construction EC type approval no.
Specifications Model-specific Data Weighing Data for Signum® Regular SIWADCP-1,-2,-3 Weighing capacity (kg) 7 16 Calibration weight (in grams) 5000 5000 Accuracy class F2 F2 35 10000 F2 65 20000 F2 Resolutions for Signum® Advanced -S Resolution >= 60,000 d -H Resolution >100,000d -I Resolution >30,000d Weighing Data Signum® SIWAEDG-3-16-S Weighing capacity (kg) 16 35 Readability (g) 0.2 0.5 Linearity (g) 0.6 1.5 Repeatability (g) 0.
SDCP: 150 238 304 360 132.5 +10 Model-specific Data Weighing Data Signum® Supreme SIWSDCP/S-1,-2,-3 ( * models not available for SIWSDCS) Non-verifiable models Standard-Version: Weighing capacity (kg) 3* 6 15* 16 3.5/16 3.5/16 168 Readability (g) 0.1 0.1 0.5 0.1 0.1/1 0.1/1 Resolution code -I -S -I -H -K -T Adjustment value 1000 5000 10,000 10,000 10,000 10,000 (in grams) Accuracy class F2 F2 F1 F1 F1 F1 7/35 0.1/1 -D 10,000 7/35 0.1/1 -P 10,000 35 0.
Dimensions Dimensions (Scale Drawings) Abmessungen (Maßskizzen) +10 132.5 +10 132.5 SIWRDCP| |SIWADCP SIWADCP| |SIWSDCP: SIWSDCP: SIWRDCP 45 45 Standard version: Standard-Version: 240 240 55 55 168 168 With integrated Akku: rechargeable battery: Mit integriertem 165 165 95 95 415 ±2±2 415 350 350 SIWSDCS: 453.5 260 31 237.5 +20 7 55 148 5 41.5 124.5 370 150 238 M 8 x1 192 30 25 SW 30 80 150.5 304 360 165.5 25 70 120 24.
Dimensions Dimensions (Scale Drawings) SIWABBP, SIWSBBP: Ø 165 ±2 85 +5 146 +5 152 Ø 35 14 205 ±2 237 ca. 500 ±5 ca. 45 206 152 194 182 194 116 ...121 85 ...90 23 182 Ø 35 ca. 500 ±5 ca.
Dimensions Dimensions (Scale Drawings) SIWSBBS: 182 117 +5 123 +5 182 34.5 215 ca. 45 205 ca.
Overview of accessories Accessories Indicator accessories: Printer and printer accessories Verifiable data printer with date, time and statistics program YDP20-0CE 5x50 m paper rolls for data printer 6906937 Replacement ink ribbon cartridge for printer 6906918 Verifiable strip/label printer with thermal printing unit, paper width up to 60 mm, with external mains unit 100-240 Volt Connection cable required Verifiable strip/label printer with thermal printing unit, paper width up to 108 mm, with extern
Overview of Accessories Item Interfaces* UNICOM: RS-232 interface module UNICOM: RS-485/422 interface module UNICOM: Analog current output interface module 0–20 mA, 4–20 mA, 0–10 V, 16 bit UNICOM: Ethernet interface module UNICOM: Dig.
Overview of Accessories Items for SIWSDCS/SIWSBBS Models Cables and Plugs Data cable for direct connection for printer YDP03-0CE Data cable for PC connection, 25-pin (5 m) Data cable for PC connection, 9-pin (5 m) Round plug for individual cable assembly Data adapter cable (round plug) for RS-232-DB25 socket for connecting standard Sartorius accessories Data cable extension (round plug/round socket) (6 m) Connection cable YDP04/12IS-0CEUV and YDP12IS-0CEUVTH to FC scales, 9-pin D-SUB plug to 12-pin round p
Declaration of Conformity Declaration of Conformity EC Conformity Mark on Sartorius Devices In 1985, the Council of the European Community approved a resolution concerning a new approach to the technical harmonization and standardization of national regulations. The organization for monitoring compliance with the directives and standards concerning the EC marking is governed in the individual EU Member States through the implementation of the EC Directives adopted by the respective national laws.
Declaration of Conformity 106 Signum 1-2-3
Declaration of Conformity Signum 1-2-3 107
Declaration of Conformity 108 Signum 1-2-3
Type-approval Certificate Signum 1-2-3 109
Plates and Markings 110 Signum 1-2-3
Plates and Markings Signum 1-2-3 111
Plates and Markings 112 Signum 1-2-3
Plates and Markings Signum 1-2-3 113
Plates and Markings 114 Signum 1-2-3
Plates and Markings Signum 1-2-3 115
Certificate 116 Signum 1-2-3
Certificate Signum 1-2-3 117
Signum 1-2-3
Signum 1-2-3 119
Keyword Index Keyword Index Page Accessories Adjustment Animal Weighing Applications Advice Automatic Data Output Average Piece Weight Updating Averaging 102 ff. 26, 43 ff. 59 2 87 51 59 Cabling Diagram Calibrate CE Mark Checkweighing Classification Cleaning Cleaning Stainless Steel Surfaces COM1 Interface: Settings Commands (Interface) Configure Data Output Connecting a Barcode Scanner Connecting the Device to AC Power Connection Options Counting 86 26, 43 ff. 105 65 ff.
General Password Appendix: General Password After selecting the “Setup” menu item a request to enter the access password “Code” will be shown on the display for 2 seconds. The place for the first character of the password flashes. Press repeatedly k, ); Press repeatedly k, ); Press repeatedly k, ); Press repeatedly k, ); Press repeatedly k, ); Press repeatedly k, ); Press repeatedly k, ); Enter password – Enter the numbers with the k key and save with the ) key.
Signum 1-2-3
Signum 1-2-3 123
Sartorius Weighing Technology GmbH Weender Landstrasse 94–108 37075 Goettingen, Germany Phone +49.551.308.0 Fax +49.551.308.3289 www.sartorius-mechatronics.com Copyright by Sartorius, Goettingen, Germany. No part of this publication may be reprinted or translated in any form or by any means without prior written permission from Sartorius. All rights reserved. The status of the information, specifications and illustrations in this manual is indicated by the date given below.